• Nenhum resultado encontrado

Displacement Measurement, Linear and Angular

N/A
N/A
Protected

Academic year: 2023

Share "Displacement Measurement, Linear and Angular"

Copied!
204
0
0

Texto

Pots are available in different tapers that determine the shape of the output function. The amplitudes of the output voltages of the secondary coils depend on the position of the core.

Table 6.1 lists some attributes inherent to pots. This chapter describes the different types of pots available, their electrical and mechanical characteristics, and practical approaches to using them for precision measurement
Table 6.1 lists some attributes inherent to pots. This chapter describes the different types of pots available, their electrical and mechanical characteristics, and practical approaches to using them for precision measurement

Piezoelectric Transducers and Sensors

An equivalent circuit must be considered to connect the transducer output voltage to the piezoelectric charge (Figure 6.43(a)). The first example is a ceramic transducer that couples the deformation of the piezoelectric sensor to the output voltage via the direct piezoelectric effect.

FIGURE 6.38 Schematic representations of the direct and converse piezoelectric effect: (a) an electric field applied to the material changes its shape; (b) a stress on the material yields a surface charge.
FIGURE 6.38 Schematic representations of the direct and converse piezoelectric effect: (a) an electric field applied to the material changes its shape; (b) a stress on the material yields a surface charge.

Plethysmography Theory

The choice of indicator used to measure volume depends on the application and may be based on temperature, color or radioactivity. The change in volume of the chest is equal to the change in volume of the chest.

Numerical Integration with Imaging

The first section is (randomly) chosen between the first coordinate of the function, and this coordinate plus the step size h. Indicated are the cutting distance h and the scanning angle α with the axis of the probe.

FIGURE 13.3 A second example of electrode positioning for impedance plethysmography, here used to estimate stroke volume: the outer two supply current while the inner two measure voltage
FIGURE 13.3 A second example of electrode positioning for impedance plethysmography, here used to estimate stroke volume: the outer two supply current while the inner two measure voltage

Indicator Dilution Methods

It is also possible to obtain other parameters; for example amplitude and phase of left ventricular ejection. Using the law of conservation of matter, the volume of the lungs can be calculated.

FIGURE 13.8 The time-concentration curve of the indicator is used to obtain the unknown volume of the fluid.
FIGURE 13.8 The time-concentration curve of the indicator is used to obtain the unknown volume of the fluid.

Water Displacement Volumetry

13.24) where Ci is the circumference of the disk at position i with assumed new radius. Consequently, due to its ease of application, the disc model method is the method of choice to measure limb volumes.

FIGURE 13.10 Illustration of the disk model to measure leg volume. The leg is divided in different sections with a fixed intersection distance the circumference of each section is obtained
FIGURE 13.10 Illustration of the disk model to measure leg volume. The leg is divided in different sections with a fixed intersection distance the circumference of each section is obtained

Equipment and Experiments

A disc height of 10 cm will be sufficient to measure the bone volume with an accuracy of at least 95%. However, it may be necessary to use a small interdisc section to obtain a higher accuracy, which may be necessary to accurately monitor the edema changes in volume.

Evaluation

One explanation is that blood filling a segment of the body lowers the impedance of that segment. The resonator frequency is inversely proportional to the square root of the volume of air in the resonator.

Introduction

Measurement of Linear Velocity

However, they give a speed relative to the transducer speed at the start of the test. The frequency of these transitions from light to dark is proportional to the speed of the moving target.

FIGURE 16.2 Velocity transducers (LVT).
FIGURE 16.2 Velocity transducers (LVT).

Velocity: Angular

After amplification, these signals are demodulated into a DC signal that is proportional to the rotation of the sensor. As shown in Figure 16.16, the magnitude of the differenced signals, a1 and a2, is divided by the distance between the two sensors, l.

FIGURE 16.7 Ac tach-generator.
FIGURE 16.7 Ac tach-generator.

Conclusion

Frequency Range • Sensitivity, Mass and Dynamic Range • Transient Response • Full Scale Range and Overload Capability • Environmental Conditions 17.13 Signal Conditioning. For practical purposes, accelerometers can also be classified as mechanical or electrical, depending on whether the restoring force or other measuring mechanism is based on mechanical properties, (eg on electrical or magnetic forces.

Accelerometer Dynamics: Frequency Response, Damping, Damping Ratio, and Linearity

However, statistical methods and probability theory can be applied to FIGURE 17.2 Logarithmic relationship between acceleration, velocity, and displacement. Therefore, equation 17.5 can be generalized by taking into account the effect of base motion.

FIGURE 17.3 Time response of a shock excitation of a single degree-of-freedom system. As the duration of the shock pulse increases, sustained oscillations get shorter in time but larger in amplitude
FIGURE 17.3 Time response of a shock excitation of a single degree-of-freedom system. As the duration of the shock pulse increases, sustained oscillations get shorter in time but larger in amplitude

Electromechanical Force-Balance (Servo) Accelerometers

A pickup mechanism detects the displacement of the coil under FIGURE 17.4 Unit step-time responses of a second-order system with various damping ratios. Displacement of the coil due to acceleration causes an electrical potential in the coil to be sensed and processed.

Figure 17.6 illustrates one form of accelerometer making use of the above principle. The coil is located within the cylindrical gap defined by a permanent magnet and a cylindrical soft iron flux return path.
Figure 17.6 illustrates one form of accelerometer making use of the above principle. The coil is located within the cylindrical gap defined by a permanent magnet and a cylindrical soft iron flux return path.

Piezoelectric Accelerometers

Under steady state conditions, motor speed is a measure of the acceleration acting on the instrument. The upper frequency of the accelerometer is limited by the resonance of the PZT crystal.

FIGURE 17.9 Frequency response of a typical piezoelectric accelerometer. Measurements are normally confined to the linear portion of the response curve
FIGURE 17.9 Frequency response of a typical piezoelectric accelerometer. Measurements are normally confined to the linear portion of the response curve

Piezoresistive Accelerometers

Differential-Capacitance Accelerometers

Denoting the gap size for zero acceleration as h0, the value of h in the presence of acceleration a can be written as:. A commonly used example of a capacitive-type accelerometer is based on a thin membrane with helical bends that provide the spring, proof mass, and movable plate required for the differential capacitor, as shown in Fig. 17.12.

FIGURE 17.11 A typical differential capacitive accelerometer. The proof mass is constrained in its null position by a spring
FIGURE 17.11 A typical differential capacitive accelerometer. The proof mass is constrained in its null position by a spring

Strain-Gage Accelerometers

Seismic Accelerometers

If the seismic instrument has a low natural frequency and a displacement sensor is used to measure the relative motion z, then the output is proportional to the displacement of the transducer case. Electrical resistivity strain gauges are also used for displacement sensing of the seismic mass as shown in Figure 17.15.

FIGURE 17.13 A typical displacement of a seismic instrument. Amplitude becomes large at low damping ratios.
FIGURE 17.13 A typical displacement of a seismic instrument. Amplitude becomes large at low damping ratios.

Inertial Types, Cantilever, and Suspended-Mass Configuration

The beam length is adjusted so that the natural frequency is equal to the frequency of the vibrating surface - hence the resonance condition obtained. Shaft speed is proportional to the square root of the local value of the acceleration.

FIGURE 17.16 A vibrating string accelerometer. A proof mass is attached to two strings of equal mass and length and supported radially by suitable bearings
FIGURE 17.16 A vibrating string accelerometer. A proof mass is attached to two strings of equal mass and length and supported radially by suitable bearings

Electrostatic Force Feedback Accelerometers

17.44) Therefore, if the bias potential V1 is held constant and the gain of the control loop is high so that changes in the gap are negligible, the acceleration becomes a linear function of the controller output voltage V2 as:. The main advantages of electrostatic accelerators include extreme mechanical simplicity, low power requirements, lack of inherent sources of hysteresis errors, zero temperature coefficients, and ease of shielding from stray fields.

Microaccelerometers

The dynamic cancellation effect of the DETF design prevents energy loss through the beam ends. The resonant frequency of the system must be much higher than any input acceleration, which limits the measurable range.

FIGURE 17.18 A double-ended tuning fork (DETF) acceleration transducer. Two beams are vibrated 180° out of phase to eliminate reaction forces at the beam ends
FIGURE 17.18 A double-ended tuning fork (DETF) acceleration transducer. Two beams are vibrated 180° out of phase to eliminate reaction forces at the beam ends

Cross-Axis Sensitivity

Selection, Full-Scale Range, and Overload Capability

The dynamic range of the accelerometer must match the high or low acceleration levels of the measured objects. Bubbles are caused by high-frequency components of the excitation near-resonance frequency that prevent the accelerometer from returning to its steady state.

FIGURE 17.19 Vectorial illustration of cross-axis sensitivity. Accelerometers may sense vibrations not only in the direction of main axis, but also perpendicular to the main axis
FIGURE 17.19 Vectorial illustration of cross-axis sensitivity. Accelerometers may sense vibrations not only in the direction of main axis, but also perpendicular to the main axis

Signal Conditioning

A typical example of a signal conditioning circuit is given in block diagram form in Figure 17.23. This chapter begins by reviewing the basic definitions of strain and the ways in which it can be measured.

FIGURE 17.20 A typical charge amplifier. The transducer charge, which is proportional to acceleration, is first converted to voltage form to be amplified
FIGURE 17.20 A typical charge amplifier. The transducer charge, which is proportional to acceleration, is first converted to voltage form to be amplified

Fundamental Definitions of Strain

When the relationship between shear stress and shear stress is linear, the relationship between shear stress and shear stress is the shear modulus G. For the Moiré technique, the stress resolution is a function of the length of the viewing area.

FIGURE 22.2 The uniaxial force shown in Figure 22.1 produces uniaxial stress in the bar
FIGURE 22.2 The uniaxial force shown in Figure 22.1 produces uniaxial stress in the bar

Principles of Operation of Strain Sensors

The output of the bridge is the difference between the voltage at points B and D. Paths ABC and ADC are voltage dividers so that VB and VD are given by equations 22.15a and b Wheatstone bridge signal conditioners are constructed with a means of "balancing" the bridge by adjusting the ratio of the resistors so that the bridge output is initially zero. The center of the circle is labeled C and the radius R. The cardinal directions correspond to zero shear stress.

FIGURE 22.5 Gage construction of a foil or wire piezoresistive gage.
FIGURE 22.5 Gage construction of a foil or wire piezoresistive gage.

General Considerations

Hooke’s Law

The aforementioned methods used for measuring forces yield a variety of designs of measuring equipment. The challenge associated with the task of measuring forces lies primarily in sensor design. The means by which the force can be converted into a workable signal form (such as electronic signals or graduated displacements).

FIGURE 23.1 When a metal is loaded in uniaxial tension ( a ) uniaxial compression ( b ), or simple shear ( c ), it will behave elastically until a critical value of normal stress or shear stress is reached.
FIGURE 23.1 When a metal is loaded in uniaxial tension ( a ) uniaxial compression ( b ), or simple shear ( c ), it will behave elastically until a critical value of normal stress or shear stress is reached.

Force Sensors

23.7) The grid configuration of resistance strain gauges made of metal foil is formed by a photoetching process. 23.12), where d is the charge sensitivity (constant for a given crystal) of the crystal in C/N. A force F causes a change in thickness.

FIGURE 23.3 Different types of load cells [2].
FIGURE 23.3 Different types of load cells [2].

Introduction

The Origins of Biopotentials

Movement of the eyeballs within the conductive contents of the skull changes the electrical potentials. Placing the electrode in the vicinity of the eyes (on either side of the eyes on the temples or above and below the eyes) picks up the potentials associated with eye movements called EOGs.

Figure 74.2 illustrates the continuum of electrophysiological signals from the (a) heart cells, (b) myocardium (the heart muscle), and (c) the body surface
Figure 74.2 illustrates the continuum of electrophysiological signals from the (a) heart cells, (b) myocardium (the heart muscle), and (c) the body surface

Biopotentials

This extended lead system allows localization of diagnostic features, such as epileptic seizures, near the electrode [18]. Similarly, saccades inherent in eye movements and eyelid blinking can cause changes in the EOG signal.

FIGURE  74.3 Schematics showing how biopotential signals are recorded from the human body
FIGURE 74.3 Schematics showing how biopotential signals are recorded from the human body

The Principles of Biopotential Measurements

Nevertheless, EOG signals can be clinically useful in acute studies of human disorders, and therefore careful acquisition of the signal followed by appropriate analysis is used to interpret the EOG potentials. A thorough effort is needed to minimize the noise and interference by improving the electrode design and placement and optimizing the amplifier circuit.

Electrodes for Biopotential Recordings

Occasionally, carbon or carbon-impregnated polymer electrodes are also used (although they are mainly used as electrical stimulation electrodes) [20]. For research applications, similar needle or wire electrodes are sometimes connected directly to the heart muscle.

The Biopotential Amplifier

This design results in the desired differential gain distributed over two stages of the amplifier. It also achieves a very high input impedance as a result of the non-inverting amplifier front.

TABLE 74.2 Distinguishing Features and Design Consideration for Biopotentials
TABLE 74.2 Distinguishing Features and Design Consideration for Biopotentials

Circuit Enhancements

Electrical isolation limits the possibility of leakage current from the instrument being used to the patient [22]. Therefore, the front of the biopotential instrument must be designed to withstand these high voltages (Figure 74.6f).

FIGURE 74.6  (continued)
FIGURE 74.6 (continued)

Measurement Practices

To eliminate RF interference, filter capacitors should be used at the front end of the amplifier as well as at various stages of the amplifier. The CMRR characteristic of the amplifier is crucial for reducing the common-mode voltage [24].

Conclusions

Electrode leads can be shielded or at least twisted together to reduce induced electromagnetic interference. Finally, the driven right arm design [27] described earlier can optionally be used to further reduce the total voltage and effective interference.

Introduction

Blood pressure values ​​vary significantly over 24 hours according to an individual's activity [1]. Basically, three factors, namely, the diameter of the arteries, the cardiac output and the condition or quantity of the blood, are mainly responsible for the level of blood pressure.

Measurement Techniques

Unfortunately, elevated blood pressure does not ensure proper tissue perfusion, and in some cases, such as certain types of shock, blood pressure may appear adequate when peripheral tissue perfusion has almost stopped. This system, however, greatly affects other organs in the body.

Indirect Blood Pressure Measurement

The AHA has published six editions of the AHA recommendations for indirect measurement of arterial blood pressure. Systolic and diastolic blood pressure were determined by cuff pressure, with the amplitude of the oscillation a fixed fraction of the maximum.

Figure 75.1 illustrates the principle of oscillometric measurement along with auscultatory measure- measure-ment
Figure 75.1 illustrates the principle of oscillometric measurement along with auscultatory measure- measure-ment

Direct Blood Pressure Measurement

Using neural network training and processing, subtle features and nonlinear relationships of the oscillometric envelope have been modeled. The sensor must be zeroed at heart level to eliminate hydrostatic errors [73].

Reproducibility, Accuracy, and Reliability Issues and Recommendations for Corrective Measures

Less often, entering a vein requires a "cutdown," direct exposure of the vein through a skin incision. Although pressure sensors can be placed at the tip of the catheter, this presents a calibration problem if it remains in place and a clot forms near the tip of the catheter, dampening the pressure signal.

Blood Pressure Impact, Challenge, and Future

Littler et al., The British Hypertension Society Protocol for the Evaluation of Blood Pressure Monitors, J. Millar-Craig, Technical and Clinical Evaluation of the Medilog ABP Non-Invasive Blood Pressure Monitor, J.

TABLE 75.7  Common Issues of Accuracy and Reliability in Blood Pressure Measurement and Recommendations of  Preventive Action or Alternative Solution in Both Indirect and Direct Measurements
TABLE 75.7 Common Issues of Accuracy and Reliability in Blood Pressure Measurement and Recommendations of Preventive Action or Alternative Solution in Both Indirect and Direct Measurements

Doppler Measurements

The depth is of course determined by the delay time and the propagation velocity in the tissue. The amplitude of the same signal is proportional to the number of moving scatterers in the tissue volume.

FIGURE 76.1 For Doppler ultrasound, the moving object shifts the received frequency.
FIGURE 76.1 For Doppler ultrasound, the moving object shifts the received frequency.

Indicator Dilution Methods

Thus, only the amount of added indicator and the time integral of the downstream concentration, assuming proper mixing, need to be known in order to calculate the flow, i.e. cardiac output. A thermal dilution catheter is placed with the injection outlet in the right atrium of the heart and with a temperature sensor in the pulmonary artery (Bronzino [6] and Weissel et al. [7]).

FIGURE 76.3 Indicator dilution measurements. The flow can be calculated if the amount of indicator and the time course of its concentration are known.
FIGURE 76.3 Indicator dilution measurements. The flow can be calculated if the amount of indicator and the time course of its concentration are known.

Plethysmography

The first reports on the successful application of the principle were published in the mid-1930s by Molitor and Kniazuk [11]. A signal modulated by the attenuation or scattering of light in the blood volume can be recorded.

FIGURE 76.4 For occlusion plethysmography, increased volume stretches the strain gage (top) and decreases the impedance (bottom).
FIGURE 76.4 For occlusion plethysmography, increased volume stretches the strain gage (top) and decreases the impedance (bottom).

Radioisotopes

If the tissue is permeated uniformly and constantly, the activity of the isotope decays monoexponentially over time. The trauma caused by the injection of the isotope into the tissue severely disturbs the flow, as shown by Holloway [17] and Sejrsen [18].

Thermal Convection Probes

The temperature difference between the two rings is a measure of the blood flow under the probe. The kilogram is the unit of mass: it is equal to the mass of the international prototype of the kilogram (3rd CGPM, (1901).

FIGURE 76.7 For thermal convection measurement of blood flow, heat dissipation increases with blood flow.
FIGURE 76.7 For thermal convection measurement of blood flow, heat dissipation increases with blood flow.

Imagem

FIGURE 6.5 The two most common rotary pot mounts are the bushing mount (a), and the servo mount (b).
FIGURE 6.38 Schematic representations of the direct and converse piezoelectric effect: (a) an electric field applied to the material changes its shape; (b) a stress on the material yields a surface charge.
FIGURE 6.43 Full (a) and simplified (b) equivalent electrical circuits of the piezoelectric sensor connected to the voltage amplifier.
FIGURE 6.44 Equivalent electrical circuit of the piezoelectric sensor connected to the charge amplifier.
+7

Referências

Documentos relacionados

Devido às alterações ocorridas nas normas da Associação Brasileira de Normas Técnicas (ABNT) e a aprovação da Resolução CEPG 01/2016, que “Dispõe sobre a