Comparison between two calibration systems for direct and attenuated radiation qualities, diagnostic

radiology level

Ana Perini

1

, Lucio Neves

1

, Margarete Guimaraes

2

, Paulo Oliveira

2

, Vitor Vivolo

1

, Teogenes Silva

2

, Linda Caldas

1

1

Radiation Metrology Center, Instituto de Pesquisas Energéticas e Nucleares, Comissão Nacional de Energia Nuclear(IPEN-CNEN/SP), Av. Prof. Lineu Prestes, 2242, Cidade

Universitária, Sao Paulo, 05508000, Brazil

2

Centro de Desenvolvimento da Tecnologia Nuclear, Comissão Nacional de Energia Nuclear (CDTN-CNEN/MG), Rua Prof. Mário Werneck, s/n – Campus da UFMG, Belo Horizonte,

30123970, Brazil lcaldas@ipen.br According to the United Nations Scientific

Committee on the Effects of Atomic Radia- tion (UNSCEAR), medical exposures con- tribute about 20% of the average annual per caput dose to the global population.

Annually there are approximately 3.1 bil- lion diagnostic medical radiological exami- nations. Therefore, it is very important to develop and maintain a quality control pro- gram for the verification of x-ray systems.

A very important step in a quality control program, at calibration laboratories, is the establishment of laboratory intercompar- isons. In this work, a comparison between the calibration laboratories of IPEN, São Paulo, and CDTN, Belo Horizonte, Brazil, both from the Brazilian Nuclear Energy Commission, was carried out, as part of the quality control program in the metro- logical network established by the project

”National Institutes of Science and Technol- ogy – Radiation Metrology in Medicine”.

The comparisons were undertaken for di- rect and attenuated diagnostic radiology beams RQR and RQA. The comparison was based on the determination of calibra- tion factors (𝑁

𝑘

) of two ionization cham- bers and of the PPV quantity. The RQR radiation qualities were compared with a Radcal

10

X5-6 ionization chamber and the RQA qualities with a Radcal RC60, both itinerant standards. The PPV quan- tity was measured for the RQR radiation qualities with a PTW diavolt non-invasive meter. The results showed a good agree- ment between both calibration laborato- ries. The results obtained showed a maxi- mum difference of 3.0% for the 𝑁

𝑘

values and of 0.77% for the PPV quantity. It is possible to conclude that both laboratories are within an acceptable range for their calibration systems, at diagnostic radio- logy level.

116 Dosimetry in Medicine and Biology

Referências

Documentos relacionados

In this work, two laboratories, the Laboratory for Calibration of Instruments of the Nuclear and Energy Research Institute (LCI at IPEN/CNEN- SP) and the Laboratory

At the present time, the Calibration Laboratory acts in the Radiation Protection, Radiation Therapy, Nuclear Medicine and Diagnostic Radiology areas, using special set-ups with

At the present time, the Calibration Laboratory is part of the Radiation Metrology Center and it acts in the Radiation Protection, Radiation Therapy, Nuclear Medicine and

pela CNEN.. This paper, besides presenting some characteristics of the spectrometer, discusses the method o f calibrating and determining the resolution o f slow neutron

The comparison between the experimental and simulated results are in good agreement for all tested radiation qualities, and the maximum difference obtained was 2.5% for the

In Tables 4 and 5 are presented the slope values, the air kerma rates and the calibration factors for the qualities of conventional diagnostic radiology,

The Comparison Exercise tested the actual measurement capacity of the laboratories performing calibration services of radiation and irradiation monitors in Brazil,

This work aims to develop a method of degradation of ion spent resins from the nuclear research reactor of Nuclear and Energy Research Institute (IPEN),