2012 International Symposium on radiation Physics - ISRP 2012 Rio de Janeiro, RJ, Brazil, October 07-12, 2012

INTERNATIONAL RADIATION PHYSICS SOCIETY -IRPS

A NEW PARALLEL PLATE GRAPHITE IONIZATION CHAMBER AS A 60Co GAMMA RADIATION REFERENCE INSTRUMENT

A. P. Perini1*, L. P. Neves1, J. M. Fernández-Varea2, V. Cassola3, R. Kramer3, H. J. Khoury3, L. V. E. Caldas1

1Instituto de Pesquisas Energéticas e Nucleares, Comissão Nacional de Energia Nuclear (IPEN-CNEN/SP); Av. Prof. Lineu Prestes, 2242, 05508-000, SP, Brazil; aperini@ipen.br

2Facultat de Física (ECM and ICC), Universitat de Barcelona; Diagonal 645, E-08028, Barcelona, Spain

3Universidade Federal de Pernambuco, Departamento de Energia Nuclear; Av. Prof. Luiz Freire 1000, 50740-540, Recife, Brazil

The accuracy and traceability of the calibration of dosimeters is of great interest to those involved in the dose delivery in radiotherapy. The calibration procedure in radiotherapy treatments is very important and a sensitive task due to the high doses delivered to the patients.

Although 60Co gamma beams are nowadays rarely utilized in radiotherapy treatments, they are still important because they are utilized for the calibration of ionization chambers and other dosimeters in terms of air kerma and absorbed dose. Generally, the air-kerma cavity standards for 60Co gamma rays are graphite cavity ionization chambers. In this work a new parallel plate graphite ionization chamber was studied to analyze its potential use as a reference instrument at the Calibration Laboratory of the IPEN. The ionization chamber was manufactured utilizing graphite and Teflon, and coaxial cables. The collecting electrode has a diameter of 42.0 mm and thickness of 2.0 mm, and the sensitive volume is 9.4 cm3. The entrance window is made of graphite and has a thickness of 4.0 mm to reach electronic equilibrium for 60Co radiation. In order to evaluate its performance in 60Co beams, it was submitted to the following tests: stability, leakage current, saturation curve, ion collection efficiency, polarity effect and angular dependence. The characterization tests were performed using a Gammatron 60Co unit. Moreover, Monte Carlo simulations were undertaken using the EGSnrc code to study the influence of the chamber components on its response. The results obtained showed that this new ionization chamber presented an excellent performance in all evaluated tests.

Referências

Documentos relacionados

Partindo para uma análise do ponto de equilíbrio considerando o custo de oportunidade e o custo de capital, teremos a seguinte modificação de valores e análise: custos fixos somados

(Lilia Diniz) No capítulo 2 busco apresentar as concepções teóricas sobre memória e seu caráter histórico quando, mesmo sonegadas, as diversas vozes dos

Tableau 1 : Récapitulatif de l’état de conservation et de dépôt des ancres selon les types de fonds Tableau 2 : Comparatif des caractéristiques des ancres de l’Amirauté et de

Se anteriormente à ação educativa as percepções, afetos e emoções que constituem os sentidos subjetivos e a subjetividade social sobre a população de rua se

Se consideró que las ovejas habian producido por lo menos un cuerpo lúteo (CL) siempre que los niveles plasmáticos de progesterona fuera n superiores a 0,5

According to that work, the influence of the wall and stem on the ionization chamber response, utilized by the PTB (parallel-plate graphite cavity ionization chamber) as refer- ence

To cite this article: Pedro Rosário, José Carlos Núñez, Luisa Trigo, Carina Guimarães, Estrella Fernández, Rebeca Cerezo, Sonia Fuentes, Marcela Orellana, América Santibáñez,

podemos observar que o estranhamento causados por (30c) e (31c) pode ser explicado através da HVC e da minimalidade-φ: a presença do clítico nos, totalmente