A new standard cylindrical graphite-walled ionization chamber for dosimetry in 60-Co beams at calibration

laboratories

Lucio Pereira Neves, Ana Paula Perini, Linda Caldas

GMR, Instituto de Pesquisas Energéticas e Nucleares, Comissão Nacional de Energia Nuclear (IPEN-CNEN/SP), Av. Prof. Lineu Prestes 2242, Cidade Universitária, São Paulo, 05508000,

Brazil lcaldas@ipen.br The use of60Co sources is largely dissemi-

nated at dosimetry laboratories, in order to calibrate the ionization chambers uti- lized for radiotherapy dosimetry, mainly in those laboratories where there is no lin- ear accelerator available. In this work, a new cylindrical ionization chamber was developed and characterized to be used as a reference dosimeter at the Calibra- tion Laboratory of the IPEN. This ioniza- tion chamber was entirely made of high- purity graphite, with a sensitive volume of 2.34 cm3 and a wall thickness of 4.0 mm, to allow electronic equilibrium for 60Co radiation. The central collecting electrode has a diameter of 2.0 mm, and is 16.0 mm long. The support of the collecting elec- trode and the stem are made of Teflon®.

The characterization tests were conducted according to the IEC 60731 standard. All tests presented results within the limits recommended by this standard: the ioniza- tion chamber achieved saturation on the

whole tested interval (−400 V to +400 V), the highest variation in the stability test was 0.23%, the ion collection efficiency was better than 99.9%, the polarity ef- fect was lower than 0.5%, the leakage cur- rent was lower than 0.5% for all tests, the response was linear, and the angular de- pendence (360 around the central axis) was lower than 0.2%. Furthermore, in order to complete the study of this new dosimeter, Monte Carlo simulations with the EGSnrc code were carried out. The wall of the ionization chamber presented an influence of 1.5% and the chamber stem 0.8% on the ionization chamber re- sponse. These results indicate a low in- fluence of the chamber components, as expected for a reference dosimeter. There- fore, the ionization chamber characterized in this work presents potential used as a reference dosimeter at calibration labora- tories.

112 Dosimetry in Medicine and Biology

Referências

Documentos relacionados

The Calibration Laboratory (LCI) of Instituto de Pesquisas Energéticas e Nucleares (IPEN) has a parallel plate ionization chamber, from Physikalish-Technische

Using the data obtained in this work the ionization chamber system installed at the LMN can be used as a reference system for activity determination of radioactive solutions... By

The aim of this work was to quantify the influence of BSR in the response of a commercial transmission ionization chamber, in the diagnostic radiology energy range, with

In this work, an extrapolation minichamber with a planar window developed at the Calibration Laboratory (LCI) of IPEN by Oliveira and Caldas (2005) was used.. The design of the

To determine whether or not the results obtained with this ionization chamber were acceptable, the IEC 60731 (IEC, 1997) recommendations, specific for ioniza- tion chambers used

In this work, a comparative study of the response stability in X-radiation beams of four monitor chambers (a cylindrical, a parallel-plate and two ring-shaped ionization chambers)

The calibration coefficients and correction factors obtained for each ionization chamber in the CT standard radiation qualities of LCI are presented in Tables 2-5, for

The objective of this work was to compare the performance of two radiation detectors in standard CT radiation beams: a homemade ionization chamber developed at