• Nenhum resultado encontrado

Primary and secondary barriers

3.1 NCRP-151 Methodology

3.1.1 Primary and secondary barriers

Chapter 3

Structural shielding design for a

megavoltage x-ray radiotherapy unit

The purpose of radiation shielding is to limit radiation exposure to members of the public and employees to an acceptable level. The NCRP-151 report presents recommendations and technical information related to the design and installation of structural shielding for megavoltage x- and gamma-ray radiotherapy facilities.

This chapter describes the methodology adopted from the NCRP-151 for the calculation of primary, secondary barriers and door for both x-rays and neutrons, including also the TADR (time averaged dose equivalent rate) and IDR (instantaneous dose-equivalent rate) calculi. Finally, there is a section dedicated to describe the bunker characteristics and environment, which will remain constant along the study.

U, use factor;

T, occupancy factor.

The transmission factor for secondary barriers (wall, ceiling, floor or other structure designed to attenuate the leakage and scattered radiations) is computed separately considering two of the most important secondary radiation sources beyond the accelerator room: the scattered radiation from the patient and the leakage radiation.

The barrier transmission factor for radiation scattered by patient, Bps, and for leakage radiation, BL, can be computed by equations 3.3 and 3.4:

Bps= P

a(θ) WpriTd2scad2sec400

F (3.3)

where:

a(θ), fraction of the primary beam absorbed dose that scatters from the patient. Values are tabulated in the NCRP-151 document (Table B-4) as a function of the photon beam energy and the scattering angle;

dsca, distance from the x-ray target to the patient (m), usually equals to 1 m;

dsec, distance from the patient (origin of the scattered radiation) to the point protected (m);

F, field area at mid-depth of the patient at 1 m from x-ray target (cm2);

400, refers the scatter fractions are normalized to those measured for a 20 x 20 cm2field size (cm2);

P, WpriandT, as previously defined.

BL= P d2L

10−3WL T (3.4)

where:

dL, distance from the isocenter to the point protected (m);

10−3, refers that leakage radiation is limited to 0.1% of the useful beam, by IEC [28];

WL, leakage workload (Gy/week);

PandT, as previously defined.

Then, the thickness of the barriers (t) regarding primary, scattered and leakage radiation, tpri, tps, tLrespec- tively, can be calculated using equation 3.5.

t=T V L1+ (n−1)T V Le (3.5)

where:

TVL1, first TVL (cm);

TVLe, equilibrium TVL (cm).

The values of the TVLs depend on the primary beam energy, shielding material, type of radiation to shield (primary, patient scattered or leakage) and scattering angle (values in Appendix A, from NCRP-151 report [9]).

The thickness of a secondary barrier (ts) is then calculated from the values oftpsandtL and applying the two-source rule. If both thicknesses are similar, the thickness of the barrier is determined by adding 1 HVL (half value layer) to the large of the two barrier thicknesses. IftpsandtL differ by a TVL or more, the larger barrier

thickness is used.

Finally, the thickness calculated with equation 3.5 should be evaluated in terms of time averaged dose equiva- lent rate (TADR) jointly with the instantaneous dose-equivalent rate (IDR).

The TADR is the barrier attenuated dose equivalent rate averaged over a specified time or period of operation, and it is proportional to IDR. According to NCRP-151 there are two periods of accelerator operation of particular interest to radiation protection: the week and the hour.

The weekly time averaged dose equivalent rate,Rw, is the TADR at a specified location averaged over a 40 h work-week and it is expressed in Sv/week. It is used to determine compliance with the shielding design goals.

For primary barriers it is given by equation 3.6:

Rw,pri=IDRWpriUpri

0 (3.6)

where:

IDR, the instantaneous dose equivalent rate (Sv/h) measured when accelerator is working at the absorbed-dose output rate ˙D0(Gy/h) at 1 m from the source and measured at 30 cm beyond the barrier.

Similarly, for secondary barriers, theRwhas contributions from both, leakage and scattered radiation, as shown in equations 3.7 and 3.8 :

Rw,s= IDRL

WL

0

+

IDRps

WpsUps

0

(3.7)

IDRps=IDRtotal − IDRL (3.8)

where:

IDRL, instantaneous dose equivalent rate measured at the secondary barrier (a point located 30 cm beyond this barrier) and in the absence of a phantom at the isocenter (Sv/h);

IDRtotal, instantaneous dose equivalent rate measured at the same point in the presence of a phantom at the isocenter (Sv/h);

IDRps, instantaneous dose equivalent rate at the same point due to the patient scattered radiation (Sv/h).

Some regulatory bodies specify a limit for the time averaged dose equivalent rate based on the integrated dose equivalent in-any-one-hour (Rh) in uncontrolled areas, which is an extra safety condition that has been followed in this work, under the condition that this value cannot exceed 0.02 mSv in-any-one-hour [48].Rh, in mSv, depends on the maximum number of treatments,Nmax, that can be performed in any-one-hour as shown in equation 3.9:

Rh=M 40

Rw, M =Nmaxh

(3.9)

where:

h, average number of treatments per hour;

40, number of working hours per week;

Mis always≥1;

Nmax, maximum number of treatments, includes the set-up time.

Width of primary barriers,win cm, will depend on the distance from the x-ray target to the outside of the barrier (d) and can be calculated by equation 3.10, where it is considered the size of the diagonal of the largest beam (usually 35x35 cm2since the corners are clipped, at 100 cm SSD) plus 30 cm to each side [9].

w= 2·30·p

352+ 352· d dsca

(3.10)

Documentos relacionados