Angular Dependence of an Intercomparison System in Low and Medium Energies of X-Rays Beams

Felícia D.G.Rocha, Oscar T.B.Alvarez and Linda V.E.Caldas

Instituto de Pesquisas Energéticas e Nucleares Comissão Nacional de Energia Nuclear Av. Prof. Lineu Prestes, 2242, São Paulo, SP, Brazil

Body of Abstract: Ionization chambers are radiation detectors widely used for X- rays measurements and quality control tests in diagnostic radiology. In radiotherapy, the ionization chambers are used for determining the absorbed dose delivered to the patient.

An ionization chamber can take many forms, but the principle of operation is similar for all chambers. It is based on the collection of charges created by the passage of radiation, with a proper voltage maintained between the wall of the chamber and the central electrode. In this work, the dependence of detector response on angle of incident radiation was evaluated for three ionization chambers used as an intercomparison standard system in low and medium x-ray beams. The ionization chambers tested are Victoreen guarded electrode with an approximately air equivalent wall, models 415, 415A and 415B. The guarded electrode construction permits direct current measurement with a minimum amount of drift. The irradiations were carried out using two different systems: a) Rigaku Denki generator, model Geigerflex, with a Philips tube model PW/2184/00 (tungsten target and beryllium window, 60 kV), and b) Pantak model HF 320 (tungsten target and beryllium window, 320 kV). All measurements were carried out in reference conditions. The variation of the response for the chamber model 415A, was verified in function of the plane of the entrance window tilted by ±5° from its reference position, perpendicular to the axis of the radiation beam, according to international recommendations. The tests were performed measuring the chamber

response, at tilt angles of 1° apart. For models 415 and 415B respectively, the chambers were rotated around their central axis from – 180° to +180° in steps of 30°. The

responses of the chambers were normalized to the reference position (0°), and the

results obtained meet the international recommendations, i.e., the limits of variation of

response are within ±1.0%.

Referências

Documentos relacionados

Figure 3 – Backscattered X-radiation effect in the transmission ionization chamber response for diagnostic radiology level RQR 3, for experimental measurements and Monte

The similar ionization chambers with different inner collecting electrode materials make up a tandem system, which consists of two dosimetric systems with different energy

The transmission chamber characteristics, studied in this work – response linearity, charge collection efficiency, polarity effect, energy dependence, leakage current and

The results obtained for the characterization tests include: saturation curve, ion collection efficiency, polarity effect, the linearity response with the X ray tube current

The established standard beams at an industrial X-ray system Pantak-Seifert were used for the characterization tests of the ionization chambers as short- and medium-term

In this work, a comparative study of the response stability in X-radiation beams of four monitor chambers (a cylindrical, a parallel-plate and two ring-shaped ionization chambers)

The objective of this work was to compare the performance of two radiation detectors in standard CT radiation beams: a homemade ionization chamber developed at

Abstract — The performance of four ionization chambers (designed for electron beams) were compared in standard X-ray beams: Advanced Markus chamber, Roos chamber and