Nuclear Instruments and Methods 179 (1981) 11-12

© North-Holland Publishing Company

A MODIFICATION TO THE ION SOURCE PULSING SYSTEM OF A PN-400 VAN DE G R A A F F ACCELERATOR

P.R.P. COELHO, L. HOLLAND and A. SGAMBATTI Jr.

Istituto de Pesquisas Energeticas e Nucleates, Sao Paulo, Brazil

Received 20 May 1980

Simple modifications to the pulsing system of a PN-400 Van de Graaff accelerator are described. An LED at ground is driven by a low power pulse generator and is coupled to a photosensitive transistor (PST) in the high tension terminal using a perspex light guide. After suitable signal amplification, the output from the PST is used to switch the ion source RF generator. Monitoring of the ion source light intensity is effected by using a second PST at ground coupled to the ion source using a second perspex light guide.

Simple modifications have been made to the PN- 400 Van de Graaff accelerator, manufactured by the High Voltage Corporation, Inc., which enable the ion source to be pulsed by a low power pulse generator operated at ground potential. A schematic of the sys- tem is shown in fig. 1.

The pulse characteristics - repetition frequency,

and mark-to-space ratio - of the pulse generator are made equal to the corresponding desired values of the n e u t r o n pulses produced by the accelerator. Pulsing of the ion source is effected by switching the grid voltage of the RF oscillator between +85 V (ON) and - 6 5 V (OFF). This switching Voltage is in turn pro- vided by a two-state voltage amplifier switched by a

CONTROL ROOM

I..?SO,LLOSOOPE jL

PULSE GENERATOR

NEUTRON PULSES

ACCELERATOR HALL

I I

"--~ • PERSPEX ROD PHOTO J

TRANSISTOIR J

(MRD '4B)l I

~ J ~ JIONBEAM J

NE-102 TRITIUM J J

8CINT TARGET I I

I I

R.F.

GENERATOR

LED J J PHOTO TRANSISTOR

~L.~ ] PERSPI~'X ROD ] j~jlMRD 14e) ' j

s

- - 4 0 0 k V

, I

~ ' 50ms SOme Fig. 1. Schematic of the pulsing system of the Van de Graaff accelerator.

AMPLIFIER

11

(2)

12 P.R.P. Coelho et al. / A modification to the ion pulsing system 'Fable 1

Time relationship between driving pulser pulses and output neutron pulses (puLser pulses, trise=6/~s, tfall = 20#s;

neutron pulses, trise and tfall <0.05 ms)

t~ (ms) tz (ms) t3 (ms) t4 (ms)

400 50 2 0.7

50 450 2.5 0.8

1 4 0.9 0.2

4 1 2.50 0.65

PULSE GENERATOR PULSES

ilr ilr I I NEUTRON PULSES ~'~

TIME

photo transistor driven in the ON and OFF states by the light emitted from a LED fixed to the grounded base plate o f the accelerator. A Perspex rod (diameter 141 ram, length 500 ram) optically couples the LED and photo transistor.

Confirmation of the ioi; source pulsing is effected using a second photo transistor optically coupled by a perspex rod to the light emission of the ion source.

The signal from this monitor transistor is amplified prior to transmission via co-axial cable some 30 m to the accelerator control room. The neutron pulses were measured with an NE102 scintillator coupled to a fast photomultiplier.

Table 1 shows the conditions under which the pulsing system was verified. It is noted that the rise and fall times of the neutron pulse (<0.05 ms) is independent of the driving pulse length, but the delay time of the neutron pulse appears to be reduced for short (--~1 ms) driving pulses.

Referências

Documentos relacionados

Você está sendo convidado a participar em uma pesquisa. O documento abaixo contém todas as informações necessárias sobre a pesquisa que está sendo realizada. Sua

a. não existir, na legislação interna do Estado de que se tratar, o devido processo legal para a proteção do direito ou direitos que se alegue tenham sido violados; b. não se

Situado no âmbito do Direito Tributário, em constante ligação com o Direito Constitucional e o Direito Processual Civil, o presente trabalho vem tratar da

Para tal tarefa, partimos, primeiramente, de dados disponibilizados pela Secretaria Geral da Mesa da Câmara dos Deputados (no sentido de sistematização dos nomes

Este projeto tem como autor o arquiteto José Bernardo Távora, sendo composto por 30 salas, 24 salas de aula para o 1.º ciclo do ensino básico, tendo capacidade para 624

1) Que o estudo se destina a mostrar a prevalência de cardiopatias congênitas por meio do oximetro de pulso ou teste do coraçãozinho. 3) Que os resultados que se desejam

Os equídeos de tração, pelas condições higiênicas sanitárias deficientes, alimentação inadequada a sua condição de trabalho e pelo acesso as vias públicas,

Para remover residuos de água, o sal foi submetido a processo de purificação vkuo (10-6mmHg) em temperatura de 150°C por 4 horas. 0 cristal foi usinado e polido nas dimensões de