A PMMA Graphite Coated Ionization Chamber for 60Co Beam Dosimetry:

Experiments and Monte Carlo Simulation A.P. Perini, L.P. Neves, L.V.E. Caldas

Instituto de Pesquisas Energéticas e Nucleares, Comissão Nacional de Energia Nuclear (IPEN/CNEN-SP), São Paulo, Brazil

Key Words: dosimetry, 60Co beams, parallel-plate ionization chamber, Monte Carlo simulation

The accuracy and traceability of the calibration of dosimeters is of great concern to those involved in the dose delivery in radiotherapy, mainly due to the high doses received by the patients. The Calibration Laboratory of the IPEN offers calibration services for radiotherapy dosimeters, and has developed a special parallel-plate ionization chamber for use as a work standard dosimeter in 60Co beams. The ionization chamber was tested in the 60Co teletherapy unit, Gammatron II S80, which is used to calibrate clinical dosimeters (radiotherapy). The irradiation conditions for all measurements in this radiation unit were a field size of 10x10 cm2 and a source-detector distance of 100 cm. The experiments followed the recommendations of the IEC 60731 standard (IEC, 2011), and the undertaken operational tests were: saturation, ion collection efficiency, polarity effect, short- and medium-term stability, leakage current, angular dependence and linearity of response. All obtained results are within the recommended limits. As the ionization chamber, evaluated in this work, presents differences in relation to the dosimeters usually employed for 60Co dosimetry (cylindrical type), Monte Carlo simulations, utilizing the PENELOPE code, were adopted to study the chamber configuration (design and materials). The simulation results showed that the maximum effect was observed for the ionization chamber stem.

The experimental and simulation results pointed out that the dosimeter evaluated in the present work presents several advantages to be employed in 60Co beam dosimetry: low- cost, easy assembling, robustness, good performance in the characterization tests and a configuration with low influence on the chamber response.

IEC, 2011; Medical electrical equipment - Dosimeters with ionization chambers as used in radiotherapy. IEC 60731. International Electrotechnical Commission, Genève.

aperini@ipen.br

Referências

Documentos relacionados

Using the data obtained in this work the ionization chamber system installed at the LMN can be used as a reference system for activity determination of radioactive solutions... By

Furthermore, owing to the geometrical differences of this ionization chamber as compared to the commercial ones, the influence of some components on its response was studied by means

In order to evaluate the performance of this ionization chamber, some characterization tests were performed: short- and long- term stability, leakage current, saturation, ion

The aim of this work was to quantify the influence of BSR in the response of a commercial transmission ionization chamber, in the diagnostic radiology energy range, with

Thus, the calibration is usually done in situ using the KAP quantity obtained from the measurements of the air kerma with a reference ionization chamber and the irradiated area on a

To determine whether or not the results obtained with this ionization chamber were acceptable, the IEC 60731 (IEC, 1997) recommendations, specific for ioniza- tion chambers used

The objective of this work was to compare the performance of two radiation detectors in standard CT radiation beams: a homemade ionization chamber developed at

Abstract — The performance of four ionization chambers (designed for electron beams) were compared in standard X-ray beams: Advanced Markus chamber, Roos chamber and