• Nenhum resultado encontrado

Efeito do torque de remoção do implante dental no osso peri-implantar = Effect of removal torque of dental implant in peri-implant bone

N/A
N/A
Protected

Academic year: 2021

Share "Efeito do torque de remoção do implante dental no osso peri-implantar = Effect of removal torque of dental implant in peri-implant bone"

Copied!
36
0
0

Texto

(1)

UNIVERSIDADE ESTADUAL DE CAMPINAS FACULDADE DE ODONTOLOGIA DE PIRACICABA

RICARDO DE OLIVEIRA SILVA

EFEITO DO TORQUE DE REMOÇÃO DO IMPLANTE

DENTAL NO OSSO PERI-IMPLANTAR

EFFECT OF REMOVAL TORQUE OF DENTAL IMPLANT

IN PERI-IMPLANT BONE

Piracicaba 2016

(2)

RICARDO DE OLIVEIRA SILVA

EFEITO DO TORQUE DE REMOÇÃO DO IMPLANTE

DENTAL NO OSSO PERI-IMPLANTAR

EFFECT OF REMOVAL TORQUE OF DENTAL IMPLANT

IN PERI-IMPLANT BONE

Tese apresentada à Faculdade de Odontologia de Piracicaba da Universidade Estadual de Campinas como parte dos requisitos exigidos para a obtenção do título de Doutor em Biologia buco-dental, na Área de Anatomia.

Thesis presented to the Piracicaba Dental School of the University of Campinas in partial fulfillment of the requirements for the degree of Doctor in Oral Biology, in Anatomy area.

Orientador: Prof. Dr. Paulo Henrique Ferreira Caria

ESTE EXEMPLAR CORRESPONDE À VERSÃO FINAL DA TESE DEFENDIDA PELO ALUNO Ricardo de Oliveira Silva, E ORIENTADA PELO PROF. DR. Paulo Henrique Ferreira Caria.

Piracicaba

2016

(3)

(4)

(5)

DEDICATÓRIA

Dedico este trabalho aos meus pais, Olício e Francisca que, com muito amor, me incentivaram a procurar ser uma pessoa melhor e espero conseguir fazer o mesmo à minha família, Maura e Bárbara, que são minhas luzes.

(6)

AGRADECIMENTOS

 Agradeço ao meu orientador prof. Dr. Paulo Henrique Ferreira Caria, pela disposição em ensinar e ajudar, além de ter me dado o melhor presente que é a sua amizade;

 À Faculdade de Odontologia de Piracicaba, ao seu diretor prof. Dr. Guilherme Elias Pessanha Henriques, ao reitor da UNICAMP prof. Dr. José Tadeu Jorge, à coordenadora geral da pós graduação prof. Dra Cínthia Tabchoury e à prof. Dra. Maria Beatriz Duarte Gavião, coordenadora do programa de biologia buco-dental;

 À minha esposa Maura por cuidar tão bem de mim e de nossa princesa Bárbara;

 Aos meus queridos irmãos, pela certeza de poder contar sempre com eles e suas respectivas famílias;

 Aos amigos e professores que me incentivaram a procurar por mais essa graduação;

 À FESB-Bragança Paulista por aceitar que realizássemos a parte experimental da pesquisa em suas dependências e seus professores Dra. Angélica, Dr. Rafael Rodrigues, Dr. Alexander Correa Borghesan, Dra. Juliana, todos os estagiários, funcionários e alunos que não mediram esforços para o trabalho fosse o melhor possível;

 À CAPES pela bolsa de estudo.

(7)

RESUMO

Objetivo: O objetivo deste estudo foi avaliar o efeito do torque de remoção (torque reverso) de implantes dentais de titânio no osso peri-implantar de minipigs. Métodos: Foram extraídos bilateralmente os dentes P1-M1 (1º, 2º e 3º pré-molares e 1º molar) de 6 minipigs. Cada animal recebeu 6 implantes de titânio Dentifix® (Ø4.1x10 mm de comprimento, com superfície STA), três em cada lado da mandíbula. No lado direito, 3 implantes permaneceram 9 meses (9M) com atividade mastigatória proporcionada por uma prótese fixa instalada por 3 meses e, no lado esquerdo, outros 3 implantes foram colocados e removidos no mesmo momento cirúrgico (IR). Todos os 36 implantes foram removidos por torque de remoção (retriever) cujos valores foram registrados e analisados estatisticamente. Os animais foram eutanasiados logo após a remoção dos implantes com o respectivo osso peri-implantar. Cada terço (cervical, médio e apical) do osso peri-implantar foi seccionado e analisado histológica e imunoistoquímica. O teste t de Student foi utilizado para identificar diferenças estatísticas nos valores entre as amostras 9M e IR. Foram apresentadas as médias e desvios padrão dos valores do torque de remoção, com 5% significância (P <0,05). Resultados: valores do torque de remoção foram maiores na situação experimental 9M do que na IR. A análise histológica apresentou osso maduro na condição experimental 9M e osso imaturo na condição IR. O torque de remoção causou pequenas fraturas e arredondamento no contorno ósseo. A análise imunoistoquímica reforçou os resultados histológicos, houve diferença estatisticamente significantes na expressão de osteocalcina nas amostras 9M e não houve diferença estatística na expressão do colágeno I em ambas as condições experimentais (P <0,05). Conclusões: Remoção de torque reverso causou fraturas microscópicas e suavização nos sulcos ósseos periimplantares mas não comprometeu a cicatrização óssea.

Palavras-chave: Torque de remoção. Osso peri-implantar. Implante dental. Minipig. Torque reverso.

(8)

ABSTRACT

Purpose: The objective of this study was to evaluate the effect of removal torque (reverse torque) of titanium implants on the peri-implant bone in minipigs. Methods: P1-M1 teeth were extracted bilaterally from six mini pigs (BR-1). Each animal received six titanium implants, three for each side of the mandible Dentifix®(External hexagon, Ø4.1x10 mm, with STA surface). On the right side of the mandible, three implants remained for nine months (9M) under masticatory activity and on the left side the other three implants were placed and immediately removed (IR). All the 36 implants were removed by reverse torque and the recorded values were statistically analyzed. Animals were euthanized right after the removal torque, whose values were recorded. Each third (cervical, medium and apical) of the periimplant bone was extracted and analyzed histologically and immunohistochemically. Student’s t-test was used to determine statistical differences in the values between the 9M and IR samples. Data were presented as means with standard deviations. The level of significance was set at 5% (P < 0.05). Results: Removal torque values were higher in 9M than in IR samples in the experimental situation. Histological characteristics of mature bone were presented in the 9M experimental condition, whereas the immature bone characteristics were presented in the IR experimental condition. Removal torque caused small fractures and rounding in the bone grooves. Immunohistochemical analysis reinforced the histological results. The Student’s t-test showed statistically significant differences in the osteocalcin expression in the 9M samples, whereas no statistically significant differences were found in the expression of collagen I in both experimental conditions (P < 0.05). Conclusions: Removal torque caused microscopic fractures and smoothening of the peri-implant bone grooves, but without compromising the bone healing.

Keywords: removal torque, peri-implant bone, dental implant, mini pigs, reverse torque.

(9)

SUMÁRIO

1 INTRODUÇÃO 10

2 ARTIGO : Effect of removal torque in peri-implant bone 13

3 CONCLUSÃO 33

REFERÊNCIAS 34

ANEXOS 35

ANEXO 1 – Certificado Comissão de Ética 35

ANEXO 2 – Comprovante de submissão de artigo 36

(10)

1 INTRODUÇÃO

Os implantes de titânio têm sido a principal escolha para substituir dentes perdidos. Tal opção tem aumentado significativamente o número de pacientes que estão sendo tratados com próteses implanto-suportadas. O número estimado de pacientes que optaram por implantes nos últimos anos é de mais de meio milhão nos Estados Unidos, cerca de 120 000 na França, 185 000 na Espanha, 410 000 na Itália e 420 000 na Alemanha (Smith e Zarb, 1989; Gonshor et al., 2011). Isto é, em parte motivado pela confiabilidade e pela alta taxa de sucesso da reabilitação oral à base de implantes, tornando esta opção previsível e segura (Esposito et al., 2005). A taxa de sucesso de implante em próteses parciais ou fixas é de 92% a 97% (Wennerberg e Albrektsson, 2011).

Por outro lado, a taxa de insucesso do implante é de cerca de 3%. Dentre as possíveis causas de insucesso dos implantes estão: infecção (peri-implantite), tensões biomecânicas exageradas e posicionamento inadequado, que exigem sua remoção (Anitua e Orive, 2012).

A exigência estética dos pacientes tem exigido mais perícia e treinamento dos dentistas. Outro fator destacado é a falta de comunicação entre o cirurgião e o protesista (Bidra, 2010) que pode resultar em um implante mal posicionado ou com angulação incorreta (Goodacre et al., 2003). Soluções para posições de implantes mal posicionados incluem segundas cirurgias ou compensação protética, procedimentos que irão aumentar o tempo e o custo do tratamento e podem comprometer o sucesso (Lee e Agar, 2006; Akkad e Richards, 2009). Extensas osteotomias e fraturas ósseas foram relatadas como resultados do mau posicionamento dos implantes (Oduncuoglu et al., 2011).

Um implante mal posicionado pode dificultar a resolução de um problema clínico. A remoção do implante de forma traumática pode comprometer ainda mais a reabilitação protética. Além disso, aumentaria o risco de lesão vascular com importantes consequências para a manutenção da vitalidade óssea local. Por isso, é importante executar a remoção do implante de forma minimamente invasiva, principalmente para conservar os tecidos ósseos e gengivais (Gerlach et al., 2013).

Diferentes técnicas foram relatadas para a remoção de implantes de titânio que

não incluem ressecção óssea, osteotomias extensas e uso de trefinas (Smith e Rose, 2010).

(11)

11

No entanto, estas técnicas comprometem a integridade dos tecidos vizinhos e podem prescindir técnicas regenerativas extensas e onerosas para restaurar a área cirúrgica. Por conseguinte, a colocação imediata de implantes ficaria comprometida, além do custo e o tempo necessário para completar o tratamento aumentariam.

A maioria das técnicas disponíveis para remover implantes mal posicionados não permite medir quantitativamente a força necessária para remover o implante. Estudos realizaram medições e avaliaram a eficácia de diferentes técnicas para de remoção de implantes mal posicionados ou falhos, mas nenhum estudo avaliou os efeitos dessas técnicas sobre o tecido ósseo (Stajčid et al., 2015). Mesmo sendo conhecido como o procedimento mais conservador para o tecido ósseo e de melhor desempenho, o torque de remoção do implante foi avaliado de diferentes formas, mas não o osso peri-implantar. Este estudo foi aprovado pelo Comitê de ética no uso animal da Universidade -CEUA / UNICAMP (Campinas, SP) (no.2730-1 / 12). Foram utilizados no experimento seis minipigs adultos machos (BR-1, São Paulo, Brasil) com aproximadamente 36 meses de idade foram e 55 kg. Os minipigs foram mantidos e adaptados no Centro Experimental da Faculdade de Medicina Veterinária (FESB-Bragança Paulista, SP), uma semana antes da cirurgia. No início do estudo, todos os animais foram submetidos a um exame físico por um veterinário e foram considerados saudáveis. Durante o período do estudo, cada minipig foi pesados e avaliadas as condições de alimentação. Os minipigs foram mantidos em baias individuais cimentadas, com água ad libitum. Doze horas antes da cirurgia, os animais permaneceram em jejum com água ad libitum. Após as cirurgias os animais foram monitorados quanto aos sinais de deiscência ou infecção do local operado e, semanalmente para avaliar a saúde geral.

Os animais foram pré-medicados antes da anestesia com midazolam (0,2mg / kg) (Medley, Sumaré, SP, Brasil) e Chlorpromazina IM (0, 1 mg / kg) (Cristália, Itapira, SP, Brasil). Um tubo endotraqueal foi usado para intubação, e uma mistura de isoflurano (Baxter Healthcare Corporation, IL, EUA) com o oxigénio numa proporção 1: 1 (5-10 ml / kg / min) foi utilizada para manter a anestesia durante a experiência. A anestesia local foi realizada com lidocaína 2% com epinefrina 12.5μg / mL (Xilocaína / Adrenalin®, Astra, Wedel, Alemanha). Após a cirurgia foi administrado pentabiótico Reinforced antibiótico

(12)

40.000 UI / kg IM (Eurofarma, Itapevi, SP) e anti-inflamatório dexametasona 3 mg / porco (MSD, São Paulo, SP, Brasil).

Procedimentos cirúrgicos para remoção de implantes

Todos os procedimentos cirúrgicos e radiográficos foram realizados pelo mesmo operador. Os dentes P1, P2, P3 e M1 (3 pré-molares e primeiro molar) foram extraídos bilateralmente da mandíbula de cada animal. Após 4 meses de cicatrização foram colocados 3 implantes hexágonos externos (EH) (Dentifix) Santa Rita do Passa Quatro, SP - Brasil) com o mesmo diâmetro e comprimento (ø4.1 x 10 mm) com a superfície STA no lado direito da mandíbula. Seis meses mais tarde, os animais receberam um implante protético para aumentar a tensão do osso (23). Três meses mais tarde, os 3 implantes com prótese foram removidos, perfazendo um total de 9 meses (9M) e do lado esquerdo da mandíbula, três implantes foram instalados e imediatamente removidos no mesmo ato cirúrgico (IV). Cada minipig recebeu 6 implantes, 3 em cada lado da mandíbula. Perfazendo um total de 36 implantes instalados. Os porcos foram anestesiados como descrito acima e todos os implantes foram cuidadosamente removidos por torque de remoção sentido anti-horário, por controlador de binário (Retriever Máximo - Belo Horizonte, MG, Brasil). Os valores do torque de remoção foram auferidos pelo aparelho mark-10 sensor de série binário universal STW (JLW Instruments, Chicago, IL, EUA). Depois, os animais anestesiados foram sacrificados com aprofundamento anestésico, as mandíbulas foram removidas e o respectivo osso peri-implante foi seccionado em pequenos blocos (10x10x6 mm).

Este estudo teve como objetivo avaliar, histológica e imunoistoquímicamente, o efeito do torque de remoção sobre o osso peri-implantar de implantes dentários instalados na mandíbula de mini pigs BR-1.

(13)

13

2 ARTIGO :

Effect of removal torque in peri-implant bone

Submetido ao periódico: Journal International of Medical Research (Anexo 1)

ABSTRACT

Purpose: The objective of this study was to evaluate the effect of removal torque (reverse torque) of titanium implants in peri-implant bone. Methods: The P1-M1 teeth were extracted bilaterally of 6 mini pigs (BR-1). Each animal received 6 titanium implants, three for each side of mandible. On the right side of mandible, 3 implants reminded 9 months (9M) under masticatory activity and on the left side, other 3 implants were placed and immediately removed (IR). All 36 implants were removed by removal torque and the recorded values were statistically analyzed. Animals were euthanized right after the removal torque and recording. Each third (cervical, medium and apical) of peri-implant bone was extracted and analyzed histological and immunohistochemically. Student’s t-test was used to determine statistical differences in the values between the 9M and IR samples. Data were presented as means with standard deviations. The level of significance was set at 5% (P < 0.05). Results: Removal torque were higher in 9M experimental situation than IR. Histological characteristics of mature bone was presented in the 9M experimental condition and immature bone characteristics was presented in the IR experimental condition. Removal torque caused small fractures and rounding in the bone grooving. Immunohistochemical analysis reinforced the histological results; student T test provided statistically significant differences to osteocalcin expression in 9M samples and no statistically significant differences expression to collagen I in both experimental conditions (P < 0.05). Conclusions: Removal torque caused microscopical fractures and smoothing in the peri-implant bone grooves but it do not compromise the bone healing ,

(14)

1 INTRODUCTION

Since the discovery of osseointegration by Branemark in Sweden in 1960, where

found that when titanium screws left undisturbed in bone, the osteocytes grow in close apposition to the titanium surfaces and provide firm anchorage. This discovery was successfully applied in dental and craniofacial reconstructive surgery in 1965. (1).(2) Dental implants became a common procedure in the modern dental treatment with long term success rates exceeding 90% reaching up to 100% (3,4) due to the development of some implant systems(5). However, the increased use of dental implants also improved the fails. The main causes of failure are incorrect position, fracture, peri-implantitis, chronic diseases and smoking (6–9)

Several studies indicated that screw loosening appeared to be one of the most common complications in dental implants once osseointegration has occurred, especially in single-tooth implant restorations(8,10,11). The incorrect position of implants can cause maxillary sinus membrane damage, pressure on the dental nerves or difficulties in prosthetic procedure as well as inconvenient esthetical problems. Esthetical requirements of patients have increased, especially for anterior teeth (12,13). Even after successful osseointegration, the implant remotion may be necessary (1,12,14,15).

To correct the wrong position or fractured implant is necessary to remove it. For this purpose, it may be used various surgical techniques such as the use of trephine, implant drills, ultrasound and others. But the use of these techniques cause great loss of peri-implant bone, what limits or prevents a new immediate rehabilitation(16,17). Alternatives to removal implants without losing or expanding alveolar bone led to Anitua & Orive (2012) to use the counter torque. Studies comparing counter torque with trephine drills to remove implants indicated better performance for the first (18,19).

The causes of implant failure are well known and described however, what happened with the peri-implant bone that can influence on the success of a reimplantation needs to be better described. With the increase of implant removals to replacement for functional or aesthetic corrections and the need to reduce alveolar bone loss(20).

(15)

15

What happens in the peri-implant bone implants removed is not reported in scientific articles. This study evaluated the peri-implant bone after his immediate removal and after 9 months of osseointegration. The aim on the present study was to evaluate the peri-implant bone after dental peri-implant removal.

2 MATERIALS AND METHODS

Animals and preparation

This study was approved by the University Animal Ethics CommitteeCEUA/UNICAMP-(Campinas, SP) (no.2730-1/12). Six adult male mini pigs (BR-1 minipigs, São Paulo, Brazil) with ~ 36 months old were and weighed ~ 55 kg used in the experiment. The mini pigs were kept in the Experimental Center of the Veterinary Faculty (FESB-Bragança Paulista, SP) and were allowed to adapt to the environment one week prior to surgeries. At the beginning of the study, all animals underwent a physical examination by a veterinarian and were found to be healthy. During the study period, the mini pigs were weighed if abnormalities in food intake were observed.(21). The identification of the animals was enabled by an marking earrings numbered. The mini pigs were kept separately in cemented stalls. Fresh water was available ad libitum. For 12 hours before surgery the animals were fasting with water libitum. The animals were inspected after the first few postoperative days for signs of wound dehiscence or infection and weekly thereafter to assess general health.

The removal torque and the histological and Immunohistochemical analysis of peri-implant bone were conducted in the mandible of the mini pigs.

The animals were premedicated to induce anesthesia with Midazolam (0,2mg/kg) (Medley, Sumaré, SP, Brasil) and Chlorpromazina IM (0, 1 mg/kg) (Cristália, Itapira, SP, Brasil). An endotracheal tube was used for intubation, and a mixture of Isofluran (Baxter Healthcare Corporation, IL, USA) with oxygen in a ratio 1:1 (5–10 mL/kg/min) were used to maintain anesthesia during the experiment. Local anesthesia was performed with Lidocaine 2% with Epinephrine 12.5μg/mL (Xylocain/Adrenalin®, Astra,Wedel,Germany). After surgery was administered IM application veterinarian

(16)

Pentabiotic Reinforced antibiotic 40.000 UI/kg (Eurofarma, Itapevi, SP) and anti-inflammatory Dexamethasone 3 mg/pig (MSD, São Paulo, SP, Brasil).

Surgical Procedures and Implants removal

The same operator performed all the surgeries and radiographic. The P1, P2, P3 and M1 teeth were extracted bilaterally of each animal. The tooth extractions were difficult in every case because the roots were divergent and usually curved distally. It was necessary to odontosection before extracting them (22). After 4 months of healing were placed 3 external hexagons implants(EH) (Dentifix. Santa Rita do Passa Quatro, SP - Brasil) with the same diameter and length (ø4.1×10 mm)with STA surface on one side of the mandible (fig. 1 and 2). The side was chosen by lot. Six months later this implant group received a prosthetic(fig. 3) to improve the bone tension (23). Three months later, the 3 implants with prosthetic were removed, totalizing 9 months (9M) and opposite side of the mandible three news implants were placed and immediately removed (IR). Each miniature pig received 6 implants, 3 on each side of the mandible. Thus, a total number of 36 implants were placed. Pigs were anesthetized as described above and all dental implants were carefully removed by a counterclockwise force (removal torque) with a torque driver (Retriever Maximus - Belo Horizonte, MG, Brasil) and the level of torque required to remove the implant from bone was recorded by mark-10 universal torque series sensor STW and removal torque were read by a force/torque indicator model BGI (JLW Instruments, Chicago, IL, USA). Afterwards, anesthetized pigs were euthanized with pentobarbital, their mandibles were cut and the respective peri-implant bone was removed in small blocks (10x10x6 mm).

(17)

17

Figure 2- Radiographs illustrating implants in the mandible of minipigs. (a) Radiography of the minipigs head with implants in both mandible sides (b) Periapical radiograph of implants position.

Figure 3- Prothesis fixed installed on the three implants.

Histology and Immunohistochemical analysis

The mandibles were sectioned into left and right segments, and each peri- implant bone was sectioned again to individualize them. Each peri-implant bone blocks were fixed in buffered formalin solution, pH 7.0, for 6 days and demineralized in 10% formic acid, dehydrated through progressing alcohol concentrations and paraffin-embedded. Paraffin blocks were sectioned at 7 µm thick mounted on poly-l-lysine coated glass slides (Sigma– Aldrich, Gillingham, UK) and processed for hematoxylin-eosin staining and for Immunohistochemical analysis. Each peri-implant bone paraffin block was longitudinal and colored with hematoxylin-eosin (H&E) to identify sites of new bone destruction or remodeling.

(18)

Immunohistochemical analysis was performed on duplicate tissue sections of peri-impant bone from each experimental specimen (9M and IR- randomly chosen). Sections were deparaffinised and rehydrated by rinsing with xylene for 10 min, industrial methylated spirit for 5 min and more 5 min in tap water. In order put out endogenous peroxidase activity, sections were incubated at room temperature in 3% hydrogen peroxide for 10 min. Two proteins were evaluated due to their sequential expression during bone healing. Collagen type I because it is expressed early in the healing process (24). Osteocalcin because is a late marker of bone formation and is expressed during mineralization by osteoblastic cells (24). To prevent non-specific protein binding was used serum-free blocking agent (DAKO, Hamburg, Germany). The sample was allowed to react for 1 h at room temperature with a primary anti-collagen I-antibody (Abcam, Cambridge, UK) and anti-osteocalcinantibody (Takara Biomedicals Europe, France). Immunohistochemical analysis was performed at different third of the peri-implant bone (cervical, medium and apical). Each third was selected at least two times per sample and analyzed. Samples images were captured then observed by means of Leica DM 4000 light microscopy (Leica Cambridge Ltd, Cambridge, UK) incorporating a Leica DFC 320 camera (Leica Cambridge Ltd) for computerized images in histological and immunohistochemistry analysis with a 40x magnification.

Image and Statistical analysis

Hematoxylin-eosin stained sections images were digitized and analyzed in order to recognize the presence of native bone tissue by the presence of osteocyte lacunaecontaining cells and the newly formed bone tissue recognized by the absence of lacunae. Also were analyzed the characteristics of peri-implant bone, presence or absence of bone fractures and the shape and contour of bone grooving resultant of the trephine action. Histological analysis was performed in images of the semi-serial slices of each peri-implant bone. They were captured by a digital camera (Samsung, South Korea) coupled to a light microscope (Zeiss, Germany) with original 200 x magnification and resolution of 600 dpi. Images around 116-80 cm were captured of each third of the peri-implant bone. Then, a digital framework of entire peri-implant bone was built by the combining three images.

(19)

19

Immunohistochemical analysis also was performed on three thirds on each sample with collagen I and osteocalcin. The same images capture and construction were made but they were measured and the value were defined by the positive staining samples, and was used to automatically analyze images of all samples that were stained under

identical conditions for both proteins and implant removals.

In the analysis of both mandible sides, the images were acquired at 200x magnification using a Nikon E600 microscope (Nikon Instruments Inc, Melville, USA. The integral optical density (IOD) of target protein was measured with Image-Pro Plus 5.0 (Media Cybernetics, Rockville, MD, USA). In the process of measurement, the values was, defined firstly by determining the positive staining of control sections , and was used to automatically analyze images of all samples that were under identical conditions

(u/pixel)(25).

Statistical analyses were performed with SPSS software (SPSS, Chicago, Ill). Student’s t-test was used to determine statistical differences in the values between the 9M and IR samples. Data were presented as means with standard deviations. The level of significance was set at 5% (P < 0.05).

3 RESULTS

Clinical observation

No remarkable complications were found during the healing period. At sacrifice, all 18 implants fixed after 9 months were considered successfully integrated at the time of the removal and none showed any mobility or signal of infection at sacrifice. There was no difference in the healing between animals who had the implants immediately removed (IR) after installation and animal whose implants were removed 9 months (9M) later of installation

Removal torque

(20)

and 2 for both experimental specimens. The removal torque values increased after 9 months, with significant differences between IR and after 9M specimens.

Table 1. Removal torque value (Ncm) of 3 implants immediate removed (IR) per animal.

Animal Mean Std. Deviation Minimum Maximum

1 98.3 5.5 92.2 103.3 2 91.6 9.1 82.1 102.5 3 105.3 8.3 100.4 115.0 4 71.6 10.5 61.2 82.2 5 78.6 5.8 72.7 83.1 6 88.6 6.6 81 93.6

Table 2. Removal torque value (Ncm) of 3 implants removed after 9 mouths (9M) per animal

Animal Mean Std. Deviation Minimum Maximum

1 150.1 30.2 122.7 184.4 2 163.3 35.1 132.4 205.3 3 175.2 15.2 153.2 204.6 4 163.6 15.4 157.3 185.1 5 153.3 15.2 146.2 174.2 6 150.3 26.4 129.2 174.6 Histological analysis

Each third of the peri-implant bone was evaluated and showed not representative difference in the bone conditions for each experimental specimens separately (9M and IR) (Fig. 4 and 5). Removal torque did not alter the characteristics of mature bone and the healing process. Thereby, did not cause significant damage in the periimplant bone. After surgical trauma, was possible to notice inflammatory process, which blood cells in the alveolar bone of IR specimens. At the 9M specimens mature bone was evident, as well as presence of fibrous connective tissue without evidence of inflammatory infiltrate. A vital bone with many osteocytic lacunae was observed on the grooving of the internal wall of peri-implant bone. Many capillaries were present, and a

(21)

21

rim of osteoblasts was observed on the bone margins. Natural inflammatory and bloody cells were visible only in IR specimens. As well as only in the IR specimens were observed small fractures and rounding in the bone grooving caused by implant trephine and removal torque. At 9M experimental condition, bone grooving presented clear contours, without rounding or fractures. In both experimental specimens there was no evidence of bone formation particularly at tissue around the peri-implant bone surface. Only in the last third (apical) was possible to identify some bone fragments, probably caused by implantation procedure.

Figure 4. Representative photomicrographs of each third of the peri-implant bone of 9M experimental condition (H&E, 40x). (A) First third (cervical third), (B) intermediate third and (C) apical third. Bone grooving with no altered contour.

Figure 5. Representative photomicrographs of each third of the peri-implant bone of IR experimental condition (H&E, 40x). (A) First third (cervical third), (B) intermediate third, and (C) apical third. Note the edges of bone grooving present rounded contour, mainly in the last third.

(22)

Immunohistochemistry analysis

Duplicate sections of peri-impant bone were obtained from each implant sample

to evaluate the percentage of stained areas in order to differentiate markers of collagen I and osteocalcin within both experimental conditions Figure 6. The highest collagen I expression values were observed at the IR experimental condition and osteocalcin expression was higher at the 9M.

Figure 6: Immunohistochemical staining of osteocalcin (A) and collagen I (C) in sections from minipigs mandible from 9M and IR osteocalcin (B) and collagen I (D). There was statistically significant differences to osteocalcin in 9M samples and no statistically significant differences to collagen I samples. Magnification: 40x (A,B) and 100x (C,D)

(23)

23

Table 3. Data showing the expression of Osteocalcin in both experimental situations 9M and IR. Osteocalcin values considered (u/pixels) (P < 0.05).

9m IR

Third Mean Std. Deviation Mean Std. Deviation

1o 190 3.6 110 10.2

2o 238* 6.1 120 10.6

3o 208* 7.2 90 11.1

* Statistically significant difference to osteocalcin expression to 9M experimental condition (P < 0.05).

Table 4. Data showing the expression of Colagen l in both experimental situations 9M and IR. Colagen I values considered (u/pixel) (P < 0.05).

9m IR

Third Mean Std. Deviation Mean Std. Deviation

1o 88.2 10.8 98.3 4.4

2o 90.5 10.2 100.2 7.2

3o 90.4 9.1 102.7 6.4

There was statistically significant difference between the two experimental specimens (9M and IR) in Immunohistochemical evaluation for osteocalcin expression. Immunohistochemical analyzes allowed to identify manifestation of osteocalcin protein in all thirds of peri-implant bone in both models evaluated, with greatest expression to model which the healing time was higher (9M). Statistical difference presented was observed especially in the middle and lower thirds. The first third presented difference, but it was not significant. The evaluation of collagen I expression did not show statistical differences. In all evaluated thirds, the presence of the protein was equivalent.

4 DISCUSSION

Dental implant revolutionized oral rehabilitation, becoming the natural teeth replacement by a titanium implant, a successful alternative to treat total or partial edentulism (14,26,27). Nowadays, dental implants are definitely a current procedure in many dental offices (3,28,29). Despite the long-term success shown by different studies (14,30) implant failure is inevitable (31–33). Since, to correct early or later failure implants is necessary to remove them, any tool available is necessary. Five different techniques to remove failing implants provided to be successful, however the counter torque technique, used on our study, is the highest predictability for insertion of another implant(17,19,34– 36)

(24)

Previous in vivo assessments of bone healing around implants presented histological observations such as bone-implant contact studies under monitored torque values(18,22,37). This study adds an extended methodology of previous investigations, because provide beyond histological analysis, Immunohistochemical analysis to assess periimplant bone behavior in a real clinical condition.

Histological analysis of early failed implants has indicated that bone overheating might be the most probable cause of failure (33,37–39) Bur-forceps, neo bur-elevator-forceps, trephine drill, and scalpel-forceps are safe implant removal techniques, however require experience and training of the operator. Counter torque technique is an easy and practice tool, because is a heating control procedure, do not requires training and can be performed by a beginner operator, so we opted to test this tool.

The clinical observations of this study showed all 18 implants fixed after 9 months were considered successfully integrated at the time of the removal and none showed any mobility (40)or signal of infection (21,33,41,42) at sacrifice.

The results of this work showed higher values of removal torque in 9M than IR specimens. It was expected since the longer healing time (9M) promotes better osseointegration than immediate implant removal. It was verify by the presence of mature bone in the peri-implant bone in the 9M specimens (4,22,43)

In order to better use a model whom reproduce the natural conditons of dental

implant in acction, has been used in this study minipigs (BR-1) (44), the nonprimate animal model most appropriate for the study of human mastication (45) and commonly used in research because suine and human share important anatomic and physiologic characteristics (46,47).

The osseointegration process is quite similar to the primary bone healing (1). After surgical procedure, there is an inflammatory process with local circulatory alteration. Afterwards, regeneration happen than bone tissue beggining to be replaced (1,48). As well as other peri-implant response happen, as the presence of collagen layer between bone and implant surfaces. The connective tissue consist in parallel collagen fibers supported by blood elements, setting the anatomical organization of collagenous

(25)

25

ligament (49,50). All histological events described above were clearly observed in all IR specimens evaluated on this study. To the 9M specimens, those events were less evident due to post-surgical time.

At the 9M specimens, alveolar bone mature bone was evident. Presence of a fibrous connective tissue with no evidence of inflammatory infiltrate. A vital bone with many osteocytic lacunae was present around the grooving implant surface. Many capillaries were present, and a rim of osteoblasts was observed on the bone margins. Natural inflammatory and bloody cells were visible only in IR experimental condition.

As all surgical procedures of our study were taken with a strict care, there was no

fracture or heating in bone tissue, which could compromise the results of this study. Long term studies indicated in histological analysis of early failed implants that, bone overheating might be the most probable cause of failure (33,37–39).

The histological analysis also presented small fractures and rounding in the bone grooving caused by implants only in the IR condition. Considering the time healing in both specimens (9M and IR), after surgical procedure some fractures and fragments were produced and, those aspects were not presented after nine months due to healing time. Removal torque caused little fracture and smooth on the peri-implant bone grooves just after installation procedure (IR) however, none considerable damage or alteration compromised the bone healing. As at 9M specimens, the bone grooving presented clear contours, without rounding or fractures, demonstrating that removal torque is not a factor of dental implant failure. Even though some bone fragments were presented in the last third (apical) just in the IR procedure, it also not compromised the bone healing.

According to Christenson R.H. (1997)(24), the bone structure, metabolism and regulation is reflected by markers of resorption, formation and/or turnover. Among the markers of bone resorption is type 1 collagen degradation and maker of bone formation: Osteocalcin. Bone formation markers derive from the osteoblastic activity, formed during the different stages of osteoblasts proliferation, differentiation and of osteoid synthesis (6,51–53) namely the bone osteocalcin, alkaline phosphatase and others makers. Osteocalcin is expressed during mineralization by osteoblastic cells. (24,54–56). Those evidences supported us to analyze the expression of bone extremes activitie: resorption

(26)

(collagen I) and formation (osteocalcin). Our immunohistochemistry results expressed the bone repair, because showed higher expression of osteocalcin at the 9M specimens. Since the titanium implants were fixed for nine months, peri-implant bone was submitted to masticatory tension(23) and that causes bone activity, stimulating osteocalcin expression, because it occurs during mineralization. Notwithstanding, collagen I expression not showed statistical difference between both experimental condition, in spite of all numerical values were higher to IR experiment. It can also be explained by the healing time evaluate. Immediate implants removal caused histological evident but has not time enough to express changes in the expression of collagen type I. The healing time was not extended because immediate removal represents a clinical situation in titanium implant procedures, when failure is detected just after its installation. The higher numerical values of collagen I expression to IR experiment condition indicate more protein activity than 9M. I also represent no removal torque influence in the healing process leading to the understanding that this does not hinder the immediate installation of a new implant in the same socket.

5 CONCLUSION

Implant removal torque should be higher to remove implants with long time installation than implants removed immediately after installation. Although, removal torque cause microscopical fractures and and smooth on the peri-implant bone grooves it does not compromised the bone healing.

6 ACKNOWLEDGEMENTS

We would like to thank Mario Perussi for supplying us with the customized system, implants and prosthetic components Dentifix®, to FESB, veterinary professors Rafael Rodrigues and Alexander Correa Borghesan and CAPES for scholarship and financial support.

(27)

27

7 REFERENCES

1. Brånemark PI. Osseointegration and its experimental background. J Prosthet Dent. 1983;50(3):399–410.

2. Chiapasco M, Gatti C. Immediate loading of dental implants placed in revascularized fibula free flaps: a clinical report on 2 consecutive patients. Int J Oral Maxillofac Implants. 2015;19(6):906–12.

3. Shemtov-Yona K, Rittel D. On the mechanical integrity of retrieved dental implants. J Mech Behav Biomed Mater [Internet]. Elsevier; 2015;49:290–9. Available from: http://linkinghub.elsevier.com/retrieve/pii/S1751616115001812

4. Paolantonio M, Dolci M, Scarano A, Archivio D, Placido G. and Histological Study in Man. J Periodontol. 2001;72(11):1560-71.

5. Esposito M, Grusovin MG, Coulthard P, Thomsen P, Worthington H V. A 5-year followup comparative analysis of the efficacy of various osseointegrated dental implant systems: a systematic review of randomized controlled clinical trials. Int J Oral Maxillofac Implants [Internet]. 2005;20(4):557–68. Available from: http://www.ncbi.nlm.nih.gov/pubmed/16161740

6. Chung H-J, Cho L, Shin J-S, Lee J, Ha I-H, Park HJ, et al. Effects of JSOG-6 on protection against bone loss in ovariectomized mice through regulation of osteoblast differentiation and osteoclast formation. BMC Complement Altern Med [Internet].

2014 Jan [cited 2015 Aug 11];14:184. Available from: http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=4066836&tool=pmce ntr ez&rendertype=abstract

7. Clementini M, Rossetti PHO, Penarrocha D, Micarelli C, Bonachela WC, Canullo L. Systemic risk factors for peri-implant bone loss: a systematic review and metaanalysis. Int J Oral Maxillofac Surg [Internet]. 2014 Mar [cited 2015 Aug 11];43(3):323–

34. Available from:

http://www.sciencedirect.com/science/article/pii/S0901502713011922

8. Goodacre CJ, Bernal G, Rungcharassaeng K, Kan JYK. Clinical complications with implants and implant prostheses. J Prosthet Dent. 2003;90(2):121–32.

9. Klinge B, Hultin M, Berglundh T. Peri-implantitis. Dent Clin North Am. 2005;49(3 SPEC. ISS.):661–76.

10. Jemt T, Pettersson P. A 3-year follow-up study on single implant treatment. J Dent. 1993;21(4):203–8.

11. Lenssen O, Barbier L, De Clercq C. Immediate functional loading of provisional implants in the reconstructed atrophic maxilla: preliminary results of a prospective study after 6 months of loading with a provisional bridge. Int J Oral Maxillofac Surg [Internet]. 2011 Sep [cited 2015 Oct 5];40(9):907–15. Available from: http://www.sciencedirect.com/science/article/pii/S090150271100213X

(28)

12. Glauser R, Sailer I, Wohlwend A, Studer S, Schibli M, Schärer P. Experimental Zirconia Abutments for Implant-Supported Single-Tooth Restorations in Esthetically Demanding Regions: 4-Year Results of a Prospective Clinical Study. Int J Prosthodont [Internet]. 2004 [cited 2015 Oct 6];17(3):285–90. Available from:

http://www.mendeley.com/catalog/experimental-zirconia-abutmentsimplantsupported-singletooth-restorations-esthetically-demanding-reg/ 13. Kohal R-J, Att W, Bächle M, Butz F. Ceramic abutments and ceramic oral implants. An update. Periodontol 2000 [Internet]. 2008 Jun [cited 2015 Oct 6];47(1):224–43. Available from: http://www.ncbi.nlm.nih.gov/pubmed/18412584

14. Adell R, Lekholm U, Rockler B, Brånemark PI. A 15-year study of osseointegrated implants in the treatment of the edentulous jaw. Int J Oral Surg. 1981;10(6):387– 416.

15. Lello GE. The Branemark Novum Protocol for Same-day Teeth: A Global Perspective. Br J Oral Maxillofac Surg [Internet]. 2002 Jun [cited 2015 Aug

5];40(3):268–9. Available from:

http://www.sciencedirect.com/science/article/pii/S0266435601907776

16. Kim Y-K, Kim B-S, Lee H-J, Hwang J-W, Yun P-Y. Surgical repositioning of an unrestorable implant using a trephine bur: a case report. Int J Periodontics Restorative Dent [Internet]. 2010 Apr [cited 2015 Aug 13];30(2):181–5. Available from: http://www.ncbi.nlm.nih.gov/pubmed/20228977

17. Li CH, Chou CT. Bone sparing implant removal without trephine via internal separation of the titanium body with a carbide bur. Int J Oral Maxillofac Surg. 2014;43(2):248–50.

18. Anitua E, Murias-Freijo A, Alkhraisat MH. Conservative implant removal for the analysis of the cause, removal torque and surface treatments of failed nonmobile dental implants. J Oral Implantol [Internet]. 2014 Dec 1 [cited 2015 Aug 20]; Available from: http://www.ncbi.nlm.nih.gov/pubmed/25436836

19. Anitua E, Orive G. A new approach for atraumatic implant explantation and immediate implant installation. Oral Surg Oral Med Oral Pathol Oral Radiol [Internet]. Elsevier;

2012;113(3):e19–25. Available from:

http://dx.doi.org/10.1016/j.tripleo.2011.06.035

20. Torsiglieri T, Raith S, Rau A, Deppe H, Hölzle F, Steiner T. Stability of edentulous, atrophic mandibles after insertion of different dental implants. A biomechanical study. J Cranio-Maxillofacial Surg [Internet]. 2015 Jun [cited 2015 Oct 6];43(5):616– 23.

Available from:

(29)

29

21. Markwardt J, Sembdner P, Lesche R, Jung R, Spekl K, Mai R, et al. Experimental findings on customized mandibular implants in Göttingen minipigs - a pilot study. Int J Surg [Internet]. 2014 Jan [cited 2016 Jan 16];12(5):60–6. Available from: http://www.sciencedirect.com/science/article/pii/S1743919113010868

22. Assenza B, Scarano A, Perrotti V, Vozza I, Quaranta A, Quaranta M, et al. Peri-implant bone reactions around immediately loaded conical Peri-implants with different prosthetic suprastructures: Histological and histomorphometrical study on minipigs. Clin Oral Investig. 2010;14(3):285–90.

23. Sasaki H, Koyama S, Yokoyama M, Yamaguchi K, Itoh M SK. Bone metabolic activity around dental implants under loading observed using bone scintigraphy. J Prosthet Dent [Internet]. 2009 Feb [cited 2015 Oct 5];101(2):127. Available from:

http://www.sciencedirect.com/science/article/pii/S002239130960007X

24. Christenson RH. Biochemical markers of bone metabolism: an overview. Clin Biochem. 1997;30(8):573–93.

25. Wei G, Hu Y, Zheng L, Huo J, Tang T, Deng F. Bone remodeling at microscrew interface near extraction site in the beagle dog mandible-histologic and immunohistochemical analyses. J Appl Oral Sci [Internet]. 2013;21(5):443–51.

Available from:

http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=3881839&tool=pmcen tr ez&rendertype=abstract

26. Chrcanovic BR, Albrektsson T, Wennerberg A. Immediately loaded non-submerged versus delayed loaded submerged dental implants: A meta-analysis. Int J Oral Maxillofac Surg [Internet]. 2015 Apr [cited 2015 Dec 3];44(4):493–506. Available from: http://www.sciencedirect.com/science/article/pii/S090150271400441X 27. Wennerberg a., Albrektsson T. Current challenges in successful rehabilitation with

oral implants. J Oral Rehabil. 2011;38(1):286–94.

28. Ayub EA, Shiratori FK, Bonfante G, Valle AL do, Abrão AM. Nobel Direct - Um novo conceito de implantes para ativação imediata - Relato de caso clínico^ipt. ImplantNews. 2007;4(1):70–5.

29. Nogueira-Filho G, Iacopino AM, Tenenbaum HC. Prognosis in implant dentistry: a system for classifying the degree of peri-implant mucosal inflammation. J Can Dent

Assoc [Internet]. 2011;77:b8. Available from:

http://www.ncbi.nlm.nih.gov/pubmed/21266099

30. Lekholm U, Gunne J, Henry P, Higuchi K, Lindén U, Bergström C, et al. Survival of the Brånemark implant in partially edentulous jaws: a 10-year prospective multicenter study. Int J Oral Maxillofac Implants. 1999;14:639–45.

31. Pabst AM, Walter C, Ehbauer S, Zwiener I, Ziebart T, Al-Nawas B, et al. Analysis of implant-failure predictors in the posterior maxilla: a retrospective study of 1395 implants. J Craniomaxillofac Surg [Internet]. 2015 Apr [cited 2016 Jan 13];43(3):414–

20. Available from:

(30)

32. Chiapasco M, Zaniboni M. Failures in jaw reconstructive surgery with autogenous onlay bone grafts for pre-implant purposes: incidence, prevention and management of complications. Oral Maxillofac Surg Clin North Am [Internet]. 2011 Feb [cited 2015

Aug 5];23(1):1–15, v. Available from:

http://www.sciencedirect.com/science/article/pii/S1042369910001305

33. Piattelli A, Scarano A, Piattelli M. Microscopical aspects of failure in osseointegrated dental implants: a report of five cases. Biomaterials [Internet]. 1996 Jun [cited 2015

Oct 5];17(12):1235–41. Available from:

http://www.sciencedirect.com/science/article/pii/0142961296849441

34. Stajčid Z, Stojčev Stajčid LJ, Kalanovid M, Đinid A, Divekar N, Rodid M, et al. Removal of dental implants : review of five different techniques. Int J Oral Maxillofac Surg [Internet]. 2015 Dec 10 [cited 2015 Dec 30];in press, . Available from: http://www.sciencedirect.com/science/article/pii/S0901502715014162 35. Albu S. The “double-barrel” approach to the removal of dental implants from the

maxillary sinus. Int J Oral Maxillofac Surg [Internet]. 2013 Dec [cited 2015 Aug

4];42(12):1529–32. Available from:

http://www.ncbi.nlm.nih.gov/pubmed/23684814

36. Smith LP, Rose T. Laser explantation of a failing endosseous dental implant. Aust Dent

J [Internet]. 2010;55(2):219–22. Available from: http://www.ncbi.nlm.nih.gov/pubmed/20604768

37. Piattelli A, Scarano A, Dalla Nora A, De Bona G, Favero GA. Microscopical features in retrieved human Branemark implants: a report of 19 cases. Biomaterials [Internet].

1998 Apr [cited 2015 Dec 30];19(7-9):643–9. Available from: http://www.sciencedirect.com/science/article/pii/S0142961297001580

38. Piattelli A, Piattelli M, Mangano C, Scarano A. A histologic evaluation of eight cases of failed dental implants: is bone overheating the most probable cause? Biomaterials [Internet]. 1998 Apr [cited 2015 Dec 30];19(7-9):683–90. Available from: http://www.sciencedirect.com/science/article/pii/S0142961297001725 39. Piattelli A, Scarano A, Piattelli M. Histologie Retrieved Dental Implants : 8 Years ’

Experience ( 1989-1996 )*. J Periodontol. 1998;178–84.

40. Grandi T, Guazzi P, Samarani R, Grandi G. Clinical outcome and bone healing of implants placed with high insertion torque: 12-month results from a multicenter controlled cohort study. Int J Oral Maxillofac Surg [Internet]. 2013 Apr [cited 2016 Jan

13];42(4):516–20. Available from:

(31)

31

41. Ata-Ali J, Ata-Ali F. Do antibiotics decrease implant failure and postoperative infections? A systematic review and meta-analysis. Int J Oral Maxillofac Surg [Internet]. 2014 Jan [cited 2016 Jan 13];43(1):68–74. Available from:

http://www.sciencedirect.com/science/article/pii/S0901502713002580

42. Aloufi F, Bissada N, Ficara A, Faddoul F, Al-Zahrani MS. Clinical assessment of periimplant tissues in patients with varying severity of chronic periodontitis. Clin Implant Dent Relat Res. 2009;11(1):37–40.

43. Aubin JE, Liu F, Malaval L, Gupta AK. Osteoblast and chondroblast differentiation. Bone [Internet]. 1995 Aug [cited 2015 Dec 30];17(2):S77–83. Available from: http://www.sciencedirect.com/science/article/pii/875632829500183E

44. Mariano M. Minisuíno (minipig) na pesquisa biomédica experimental . O Minipig br1. Acta Cirúrgica Bras. 2003;18(5):387–91.

45. Herring SW. How can animal models answer clinical questions? Futur Orthod. 1998;89–96.

46. Swindle MM, Smith AC, Laber-laird K, Dungan L. Farm Animals in Biomedical Research — Part One Swine in Biomedical Research : Management and Models. 1994;36(1).

47. Ruehe B, Niehues S, Heberer S, Nelson K. Miniature pigs as an animal model for implant research: bone regeneration in critical-size defects. Oral Surg Oral Med Oral

Pathol Oral Radiol Endod [Internet]. 2009 Nov [cited 2015 Aug 5];108(5):699–706.

Available from:

http://www.sciencedirect.com/science/article/pii/S1079210409004363

48. McKinney RV Jr,Steflick DE, Koth DL SB. The scientific basis for dental implant therapy. J Dent Educ. 1988;52(12):696–705.

49. McKinney RV Jr KD. The single-crystal sapphire endosteal dental implant: material characteristics and 18-month experimental animal trials. J Prosthet Dent. 1982;47(1):69–84.

50. McKinney RV Jr, Steflik DE KD. Histological results from a comparative endosteal dental implant study. J Dent Res. 1987;66(186).

51. Aubin JE. Principles of Bone Biology [Internet]. Principles of Bone Biology. Elsevier; 2008 [cited 2015 Dec 30]. 85-107 p. Available from: http://www.sciencedirect.com/science/article/pii/B9780123738844000264

52. Cremers S GP and SM. Biochemical Markers of Bone Metabolism. Princ Bone Biol 3rd ed. 2008;press:1857–81.

53. Seibel MJ. Molecular markers of bone turnover: biochemical, technical and analytical aspects. Osteoporos Int [Internet]. 2000;11 Suppl 6(suppl 6):S18–29. Available from:

(32)

54. Verbicaro T, Giovanini AF, Zielak JC, Baratto Filho F, de Araujo MR, Deliberador TM. Osteocalcin immunohistochemical expression during repair of critical-sized bone defects treated with subcutaneous adipose tissue in rat and rabbit animal model. Braz

Dent J [Internet]. 2013;24(6):559–64. Available from: http://www.ncbi.nlm.nih.gov/pubmed/24474349

55. Campos JM de, Prati AJ, Cirano FR, Pimentel SP, Pastore GP, Pecorari VG, et al. Smoking modulates gene expression of type-1 collagen, bone sialoprotein and osteocalcin in human alveolar bone. J Oral Maxillofac Surg [Internet]. 2015 Jun [cited

2015 Jun 29]; Available from:

http://www.sciencedirect.com/science/article/pii/S0278239115009039

56. Yun H-M, Ahn S-J, Park K-R, Kim M-J, Kim J-J, Jin G-Z, et al. Magnetic nanocomposite scaffolds combined with static magnetic field in the stimulation of osteoblastic differentiation and bone formation. Biomaterials [Internet]. 2016 Jan 26 [cited 2016

Feb 10];85:88–98. Available from:

http://www.sciencedirect.com/science/article/pii/S0142961216000491

(33)

33

3 CONCLUSÃO

Embora o torque de remoção cause suaves fraturas microscópicas no osso contorno do peri-implantar, estas ranhuras não comprometeram a cicatrização óssea.

Para a remoção de implantes com maior tempo de osseointegração, o torque de remoção deve ser maior do que os implantes removidos imediatamente após a instalação.

(34)

REFERÊNCIAS

Akkad S, Richards M. Solutions for severely angulated implants in the mandibular overdenture: a clinical report. J Prosthodont. 2009 Jun;18(4):342-7. doi:

10.1111/j.1532849X.2008.00400.x

Anitua E, Orive G. A new approach for atraumatic implant explantation and immediate implant installation. Oral Surg Oral Med Oral Pathol Oral Radiol . 2012 Mar;113(3):e19-25. doi: 10.1016/j.tripleo.2011.06.035.

Bidra AS. Surgical and prosthodontic consequences of inadequate treatment planning for fixed implant-supported prosthesis in the edentulous mandible. J Oral Maxillofac Surg. 2010 Oct;68(10):2528-36. doi: 10.1016/j.joms.2010.05.054.

Esposito M, Grusovin MG, Coulthard P, Thomsen P, Worthington H V. A 5-year follow-up comparative analysis of the efficacy of various osseointegrated dental implant systems: a systematic review of randomized controlled clinical trials. Int J Oral Maxillofac Implants. 2005 Jul-Aug;20(4):557-68.

Gerlach RC, Dixon DR, Goksel T, Castle JT, Henry WA. Case presentation of florid

cementoosseous dysplasia with concomitant cemento-ossifying fibroma discovered during implant explantation. Oral Surg Oral Med Oral Pathol Oral Radiol. 2013 Mar;115(3):e44-52. doi:

10.1016/j.oooo.2012.05.020.

(35)

35

ANEXOS

ANEXO 1 – Certificado Comissão de Ética

(36)

ANEXO 2 – Comprovante de submissão de artigo

Referências

Documentos relacionados

4.3 - RELAÇÕES ENTRE AS PRÁTICAS DE MARKETING E A PERFORMANCE DE EXPORTAÇÃO A análise de correlação entre as práticas de marketing e a performance de exportação teve como

This study demonstrated that higher insertion torque values in a conical internal connection increase the it (contact) of the implant-abutment interface... Keywords: Dental

Many factors may affect the quality of implant insertion, including bone density. When high values of torque are applied during EH implant surgical placement, the rotational

Thus, considering the relevance of evaluating peri- implant health for long-term success of implants, the aim of this study was to assess the peri-implant bone loss and success

The aim of this systematic review and meta-analysis was to compare the peri-implant vertical bone loss of immediate loading of implant crowns using the one abutment at one time

A primeira sugestão tem a ver com a continuidade do estudo, oficina de formação, Despertar para o ensino prático e experimental do ensino das ciências, mas

study was to evaluate peri-implant bone loss (PIBL) and stability around submerged and non-submerged dental implants in T2DM and NT2DM

Thus, the objective of this research study was to evaluate the correlation between radiomorphometric indexes of bone density, implant stability quotient and torque