Nuclear Instruments and Methods in Physics Research A 378 (1996) 624-626

NUCLEAR INSTRUMENTS

a METHODS IN PHVSICS

ELSEVIER Letter to the Editor -EIEY

Comparison of two methods for relocation of multichannel spectra

0. Helenea’*, M. Morallab

“Institute de Fisica da Universidade de Srio Paulo. CP 66318, CEP 0.5389-970 .%o Paula, SP, Brazil

“Institute de Pesquisas EnergPticas e Nucleares, CP 11049. CEP 0.5422-970 .%o Paula, SP, Brazil Received I I December 1995; revised form received 8 April 1996

Abstract

Two recently published methods of multichannel data relocation have been compared. One of them is deterministic and does not preserve the typical Poisson distribution of the data. The other one concerns a stochastic procedure that conserves the statistical fluctuation of the original data. It is shown that usual statistical tests are useless for the spectra relocated with the deterministic method.

Relocation of multichannel spectra is necessary in order to compare or add different spectra or to correct for effects of shifts and gain variations due to instabilities of the apparatus. Commonly, the new spectra produced in the relocation process must be analyzed in the same way as the original ones, by employing usual fitting procedures.

ai, =LY’II, +P’n,#m,, (2)

Recently, two methods of relocation were suggested [ 1,2]. The common point of both methods is the fit of polynomial functions to the data distribution before the redistribution of the counts among the new channels.

However, one of the methods adopts a deterministic procedure in order to distribute the data of the original spectrum [I], while the other adopts a stochastic procedure [2]. The main purpose of this paper is to compare both methods.

less than the expected value for Poisson distributed data.

The stochastic procedure attributes a number of counts m, to channel 1’ by drawing a binomial random number n : from n, with probability (I plus a binomial random number ni from nz with probability p:

m, =n{ +ni.

where

(3)

Fig. 1 shows the counts n,, n,, n3,. . . in successive channels of a spectrum to he relocated. Suppose that these counts are to be distributed among the new channels. For example the new channel 1’ must receive a fraction a of the counts in channel 1 and a fraction p of the counts in channel 2; the new channel 2’ must receive a fraction 1 - /_? of the counts in channel 2 and a fraction y of the counts in channel 3, etc., where (Y, /3, -y, etc. are calculated from the fit of polynomials to the original spectrum distribution, and integrating the fitted function over the new channels. The deterministic procedure implies that

9,,.,w = cn n,!

, - n;)!n;! (Y”i(l

ORIGINAL SPECTRUM

ch. counts

a))II+ )

(4)

RELOCATED SPECTRUM

counts ch.

ml =(~n, +&, mZ =(l -P)n,+yn,. (1)

The variances of the new counts generated with this procedure do not obey Poisson distributions. For example, the variance of m, is

2 n2

- 2’

3 n3

y m2

--f.!q

m3 3’

*Corresponding author. Fax +55 11 814 0503, e-mail Fig. 1. Scheme of the distribution of multichannel events per-

otaviano@if.usp.br. formed by a deterministic relocation method.

0168~9002/96/$15.00 Copyright 0 1996 Elsevier Science B.V. All rights reserved

PI1 SO168-9002(96)00523-2

(2)

0. Helene, M. Moralles I Nucl. Instr. and Meth. in Phys. Res. A 378 (1996) 624-626 625

P,,,.,(fll) =

(n _

nz!

n;)!n,!

P”‘(I - p)“z-“i,

2

The contents of the other relocated channels are pro- duced in a similar way. For the new channel 2’, the mZ counts are obtained by

m, = (nz - n;) + 11;. (6)

where nl is drawn from the binomial distribution P,,,,,(n:).

As shown in Ref. [2], this procedure assures that the counts of the new channels do obey Poisson distributions.

Fig. 2a shows a simulated Gaussian peak in a multichan- nel spectrum which was relocated by 0.5 channel using both the deterministic (Fig. 2b) and stochastic (Fig. 2c) procedures. Results of the fit of a Gaussian function to these peaks, using the last squares method, are also shown in the inset. The fits were performed supposing that the statistical fluctuations are given by the square root of the number of counts and neglecting covariance between counts in different channels, as usual. It should be ob- served that the fitted areas shown in Figs. 2b and 2c agree with the area of the original spectrum of Fig. 2a, since both

4L_.__l 0: : : : : l,_,..Y., : : : I- 1200 POS. 49.50(3) h 2

x M P(2) AREA 5021 (50) 0.56 95 %

(W

1200 -POS.

AREA Boo-

30 40 50 7c

CHANNEL

1

Fig. 2. A single peak fitted with a Gaussian curve plus linear background for the original spectrum (a). for the spectrum relocated by using the method of Ref. [l] (b), and for the spectrum relocated by using the method of Ref. [2] (c).

methods preserve the total counts. However, the deter- ministic relocation gives a very small chi-squared value and, consequently, a very high probability. This occurs due to the fact that the statistic fluctuation of the deterministi- tally relocated spectrum is less than the corresponding Poisson distribution. Thus, the deterministic procedure nullifies any usual statistical test of the fit. By comparing the chi-square tests shown in Figs. 2a and 2c one sees that the original and stochastically relocated spectra present very close results.

Fig. 3a shows a region of the gamma-ray spectrum of

‘“Ba taken with an HPGe detector. showing the strong peak with 356 keV and the small background 352 keV gamma-ray from ?14Pb. Results for a Gaussian function fit with a second degree polynomial for the background are shown in the inset. The relocated spectra with the de- terministically and stochastically procedures are shown in Figs. 3b and 3c, respectively. As can be seen, the presence of two peaks is very clear in all spectra but the chi-square

I ’ I

400 - POS. 1154.M (5) AREA 1750(47) ZQO-

4OrJ - POS. 1153.70 (5) AREA 1754 (47) m-

2 2 200- 0

400 - POS. 1153.72 (5) AREA 1754 (48) 300-

1125 1140 1155 1170

CHANNEL

Fig. 3. The 356 keV (“‘Ba) and 352 keV (‘14Pb) gamma-rays

fitted with a Gaussian curve plus a second degree polynomial

background for the original spectrum (a), for the spectrum

relocated by using the method of Ref. [l] (b). and for the spectrum

relocated by using the method of Ref. [2] (c).

(3)

626 0. Helene, M. h4oralle.v I Nucl. ktr. and Meth. in Phw Res. A 378 (1996) 624-626

400 t POS. 1 1140.8 (5) lY 2 rrcl wr

AREA 1 90 (20) 0.81 88

300 PO.8 2 1153.70 (5) n

2 I AREA 2 1785(48)

52ctC 8

0 wc

I

1125 1140 1155 1171

:2)-

$6.

a)

u,

*)

%,

9 24

3 CHANNEL

Fig. 4. The 356 keV (“‘Ba) and 352 keV (‘“Pb) gamma-rays fitted with two Gaussian curve plus a second degree polynomial background for the original spectrum (a), for the spectrum relocated by using the method of Ref. [I] (b). and for the spectrum relocated by using the method of Ref. [2] (c).

test applied to the deterministically relocated spectrum is compatible with a single Gaussian adjust, while for the original and stochastically relocated spectra the same test discards the single peak hypothesis. Figs. 4a-4c show the same spectra of Fig. 3. but fitted with two peaks. Now the chi-square tests for the original and stochastically relocated spectra indicate that the inclusion of two peaks is neces- sary to obtain satisfactory confidence in the fit. On the other hand, the tit of the deterministically relocated spectrum, that was already satisfactory with a single peak, gives now a very low and unrealistic chi-squared value.

The stochastic procedure has been used as a useful tool to relocate spectra obtained in experiments that need long time for the data acquisition [3] and no problem has been detected when employed within the limits described in Ref. 121.

Acknowledgement

This work was partially supported by CNPq. Brazil.

References

111 I21

131

L. Loska, Appl. Radial. Isot. 46 ( 1995) 949.

S.P. Tsai, E.R. Mucciolo and 0. Helene, Nucl. Instr. and Meth. A 345 (1994) 538.

M. Moralles, P.R. Pascholati, V.R. Vanin and 0. Helene, Appl.

Radiat. Isot. 46 (1995) 133.

Referências

Documentos relacionados

(2009) identificam três grupos de clientes que deverão ser alvo de estratégias diferenciadoras: os clientes mais valiosos, os clientes com maior potencial e os clientes

Esta carta visa explicar e exemplificar ao público um método de cálculo do Índice de Severidade de Doenças para determinação dos limites de tolerância para rizoctoniose,

2 e intitulado “Plantas úteis nativas do Brasil na obra dos naturalistas”, enviadas ao portal da Associação Brasileira de Horticultura (ABH) e também postadas em um grupo

2 This study shows a case of sepsis with pulmonary focus associated to HPeV infec- tion and should serve as a warning to the introduction of this investigation in the routine care

Embora esta dissertação tenha como objeto de estudo o Marketing desenvolvido pela CESP no se- tor de energia elétrica, cabe ressaltar que, faz parte da política

Apesar de não se ter obtido relação entre a concentração de mercúrio e o IMC e/ou índice cintura-anca, cerca de 39% dos participantes deste estudo apresentaram

Este trabalho tem por objetivo utilizar o tratamento eletrolítico operando em batelada com recirculação como um meio de desinfecção de água de abastecimento, empregando um reator

o Instituto Alana, no intuito de contribuir para a efetivação e proteção dos direitos da criança e do adolescente, vem, respeitosamente, por meio de seu