• Nenhum resultado encontrado

Sequential injection system for simultaneous determination of chloride and iodide by a Gran’s plot method

N/A
N/A
Protected

Academic year: 2023

Share "Sequential injection system for simultaneous determination of chloride and iodide by a Gran’s plot method"

Copied!
6
0
0

Texto

(1)

Sequential injection system for simultaneous determination of chloride and iodide by a Gran’s plot method

Alberto N. Araújo

a

, Maria Conceição B.S.M. Montenegro

a,

, Lucie Kousalová

b

, Hana Sklenáˇrová

b

, Petr Solich

b

, Ricardo Perez Olmos

c

aCEQUP/Department of Physical Chemistry, Faculty of Pharmacy, University of Porto, Rua An´ıbal Cunha 164, 4050 Porto, Portugal

bDepartment of Analytical Chemistry, Faculty of Pharmacy, Charles University, Heyrovskeho 1203, 500 05 Hradec Kralove, Czech Republic

cDepartment of Analytical Chemistry, Escuela Universitária de Ingenieria Tecnica Industrial, Plaza de la Castilla 3, 48012 Bilbao, Spain Received 26 July 2002; received in revised form 22 November 2002; accepted 20 January 2003

Abstract

This paper describes an automated set-up based on the sequential injection analysis (SIA) concept with potentiometric detection for the determination of chloride and iodide at low concentrations. The assessment of both ion concentrations is accomplished by titration with silver ions using the Gran’s plot approach. The proposed procedure enables chloride and iodide to be determined simultaneously in the range 6.0×106to 1.0×104mol l1if a minimum silver concentration in the volumetric solution is chosen (5×105mol l1). Conventional titrations based on the first derivative evaluation of the equivalence points applied to the same solutions fail, especially for chloride determinations.

Although, a comparison of results obtained for higher analyte concentrations shows that they have similar accuracy and precision.

© 2003 Elsevier Science B.V. All rights reserved.

Keywords: Potentiometric titration; Gran’s plot; Simultaneous determinations; Iodide; Chloride; Sequential injection analysis

1. Introduction

The field of titration procedures is the oldest yet an ac- curate tool to perform analyte determinations in samples.

However the demands for time saving and scale-down in order to reduce reagent consumption and analysis costs have encouraged the development of different automated titration procedures. In this context, flow-based procedures contributed to fulfil these goals. The titration procedure has been adapted to flow technology since Bladell and Laessig [1–3] proposed a system in which a continuous or variable addition of the analyte to a constant flow of the volumetric solution was monitored by means of a potentiometric sensor until the correct analyte flow rate enabled the stoichiomet- ric ratio. The use of a mixing chamber allowed complete homogenisation between both solutions and a prior calibra- tion procedure was necessary to correct results biased by drifts in the flow conditions.

Corresponding author. Tel.:+351-2208-7132; fax:+351-2200-3977.

E-mail address: mcbranco@mail.ff.up.pt (M.C.B.S.M. Montenegro).

With the advent of flow injection analysis[4], several ten- tative procedures were proposed in order to perform titra- tions under controlled dispersion conditions. In the more common approach, the logarithm dependence of two pseudo end-points found in the front and tail of the gradient created by injection of sample in the flow stream is used to establish a relationship between the time difference between the two events for several calibration solutions and the correspond- ing concentration.

Korn et al. proposed a flow set-up based on a binary sampling concept[5]where the analyte and volumetric so- lutions were aspirated through the same pumping channel and mixed in a non-segmented stream. The volumetric ra- tio between the analyte and volumetric solutions necessary to achieve the titration end-point was not dependent on the current hydrodynamic conditions and thus calibration pro- cedures, which are not involved in conventional titrations were also avoided.

In the present work this concept was used in a poten- tiometric multi-task flow system based on the sequential injection analysis (SIA) concept [6] to provide auto- mated calibration and on-line preparation of testing solu- tions. Gran’s plots enable linearisation of data obtained in

0003-2670/$ – see front matter © 2003 Elsevier Science B.V. All rights reserved.

doi:10.1016/S0003-2670(03)00096-5

(2)

potentiometric titrations and thus easily and precisely to lo- cate equivalence points of titrations. The theory associated with these plots is straightforward[7].

The response of an ion-selective electrode to a single charged cation, X, in solutions free from interferents, may be represented by the Nernst equation:

E=E0+SlogaX

where S is the slope of the electrode response plot and aX

is the activity of the monitored ion in the sample solution.

Rearrangement of this equation gives:

antilog

E

S

=constant(aX)

where constant is antilog(E0/S). Hence, antilog(E/S) could be plotted against the volume of the analyte added to give a linear plot with an intercept on the analyte volume axis aX=0 corresponding to the equivalence point if correction for the volume of analyte added is made. Important advan- tages of Gran’s plots result from the use of several exper- imental data obtained farthest from the equivalence point, which are insensitive to interferences that produce a sepa- ration between end-point and equivalence point. They also save time in producing reliable results even when performed under flow conditions[8].

Potentiometric determinations of chloride or iodide in flow conditions have previously been described[9–14]but could not discriminate between more than one halogen species. In this work it is proved that if more than one analyte is present in the sample to be titrated and both an- alytes are present at low concentrations, the Gran’s plots approach could provide good analytical results, even when conventional first derivative titrations fail. However, the an- alytical ranges of the analytes and the volumetric solution concentrations should be carefully selected.

2. Experimental

2.1. Reagents and solutions

All solutions were prepared with distilled, deionised wa- ter (conductivity <0.1␮S cm1). Analytical grade chem- icals were used without further purification. Potassium iodide, sodium chloride, potassium nitrate and silver nitrate (Merck) were used throughout. The iodide stock solution of 0.1 mol l1 was prepared from solid potassium iodide previously dried at 110C for 4 h, weighing the solid and dissolving 16.6010 g in 1000.00 ml of water. The chlo- ride stock solution of 0.1 mol l1 was prepared from solid sodium chloride previously dried at 250C for 1 h, weigh- ing the solid and dissolving 5.8440 g in 1000.00 ml of water.

Less concentrated solutions were obtained by subsequent dilution of the stock solutions with 0.1 mol l1 potassium nitrate.

A 0.1 mol l1 silver nitrate stock solution for the po- tentiometric determination was prepared from Titrisol vials (Merck, ref. 1.09990) by careful dilution with water. The less concentrated solutions ranging from 5.0×106to 5.0× 103mol l1 were further diluted with 0.1 mol l1 potas- sium nitrate.

To determine chloride and iodide ions with potentiometric detection a carrier solution of 0.1 mol l1potassium nitrate was used. A concentration of 1.0×105mol l1 of silver ions was maintained in the carrier solutions throughout calibrations of electrodes in order to minimise the baseline drift. Aluminium nitrate stock solution of 1×102mol l1 was prepared by weighing the solid aluminium nitrate and dissolving it in water. Concentrations ranging from 5×106 to 104mol l1 aluminium nitrate in the sample were ob- tained by adding the stock solution to the prepared sample solution in order to evaluate and avoid co-precipitation.

2.2. Electrode preparation

Tubular silver-selective electrodes with a homogeneous crystalline membrane and without internal reference so- lution were constructed as previously described [15]. The electrode membranes were maintained in contact with 1.0×103mol l1silver solution for about 12 h. The elec- trodes were kept in contact with the same solution after they were polished.

This sensor module was firmly attached to the set-up, inserted into an acrylic rectangular block and screwed on tightly. That allowed the insertion of the electrode into the SIA manifold. The quality of the joint between the flanges of the tubes connecting it to the set-up and the sensor mod- ule was ensured by a shallow circular depression with sim- ilar diameter as that of the flanges which were slotted into it. Hence, the approximately 0.4 mm long sensor surface of the tubular detectors in contact with the solutions had the same diameter as that of the SIA manifold tubes. From time to time, the membrane of the sensor modules was polished introducing damp cotton and 3␮m aluminium oxide pow- der (Buehler, 40-6603-030-016) into the channel. Whenever the detectors were not used for prolonged time periods or showed some loss of their operating characteristics, specially their response rate, they were re-polished and re-conditioned in the aforementioned manner.

2.3. Apparatus

The proposed system used for the simultaneous deter- mination of chloride and iodide by SIA is depicted in Fig. 1. It comprised a Gilson Minipuls 3 peristaltic pump (Villiers-le-Bel, France) as the propulsion device equipped with a PVC pumping tube (i.d. 1.60 mm) of the same brand and a VICI CheminertTMC25-3118 E model eight-port se- lection valve (Valco Instruments, Houston). All tubular flow paths were made of 0.8 mm i.d. PTFE tubing. A holding coil (HC) placed between peristaltic pump (P) and selection

(3)

Fig. 1. Schematic view of sequential injection (SIA) set-up developed; P, peristaltic pump; SoV, solenoid valve; HC, holding coil; SeV, selection valve; GE, grounding electrode; IE, indicator electrode; Ref, reference electrode; W, waste.

valve (SeV) was 350 cm long and was coiled over a plastic net. Port 3 of the SeV was connected to a NResearch 161 T031 (Stow) three-way solenoid valve (SoV) by means of a 9 cm long tube.

A potentiometric detection unit was connected to port 8 with a 55 cm long tube. The Ag(I)/Ag2S tubular electrode with a homogeneous crystalline membrane sensitive to silver ion, constructed according to Ferreira et al.[10], was used as the potentiometric detector with a AgCl/Ag double-junction reference electrode from Russell, model 90-0029, containing 10% potassium nitrate solution in the outer compartment.

Potentiometric measurements were made with a Crison Model 2002 digital voltmeter (±0.1 mV sensitivity) and titration curves recorded with a Kipp and Zonen model BD 111 chart recorder. The other inlets of SeV and SoV2were fitted with 18 cm long PTFE tubing. SoV1 was positioned between the peristaltic pump and the HC in order to min- imise the inertial effect of the fluids and to discard pulses which affected the reproducibility of the volumes aspirated into the system. This valve enabled the flow of the fluids inside the HC to be synchronised with the peristaltic pump movement[16].

Several laboratory-made devices, namely joints, the sup- port of the reference electrode and the earth connection contact (ground electrode) as well as the support of the silver-selective tubular electrode (construction described elsewhere [17,18]) were also used. The rotation speed of the peristaltic pump, the rotor position for the SeV and SoV on/off switching were controlled through a PCL-711 Ad-

Table 1

System control parameters used for potentiometric measurements of externally prepared solutions and on-line mixing of solutions in various ratios

Step Task SeV SoV state Volume (l) Flow rate

(ml min1)

Time (s) Flow direction (F/R)a Electrode calibration

1 Sampling 4 927 2.78 20 R

2 Detection 8 6068 2.60 140 F

Potentiometric titration

3 Sampling 3 On/off (5–75%) 927 2.78 20 R

4 Detection 8 6068 2.60 140 F

5 HC cleaning 5 607 2.60 14 F

a Forward (F); reverse (R).

vantech interface card and laboratory-made software was developed in Quick Basic 4.5 language.

An automated system for conventional potentiomet- ric titration comprised a Crison 2031 burette and Crison 2002 pH-meter, both computer controlled. An AgCl/Ag double-junction reference electrode from Russell, model 90-0029, containing 10% potassium nitrate solution in the outer compartment was used as a reference electrode and a silver-sensitive electrode of conventional shape[15] was applied as an indicator electrode.

2.4. Procedures

Initially, the holding coil and transmission line between SeV and the detection system were filled with the carrier so- lution, selecting port 8 of SeV and positioning the peristaltic pump in the propulsion mode. This configuration was kept until a stable detector signal was achieved. Afterwards, the flow direction was reversed and port 4 of SeV was selected in order to fill the respective access channel with the exter- nally prepared calibration solution. Selection of SeV port 3 and activation of SoV2enabled filling of the inlet chan- nel with silver nitrate in order to titrate and to estimate the electrode slope.

When port 3 was selected and SoV2 was inactive, the channel of the passive port was filled either with the sample solution to enable the potentiometric titration, or with the carrier solution to promote the calibration of the potentio- metric detection unit. After the solutions reached the HC, they were discharged to waste by selecting SeV port 5 and driving the pump in the forward mode.

The analytical cycle established for the potentiometric determination of chloride and iodide comprised five steps (Table 1). By sequential performance of the first two steps four times, the calibration of the electrode becomes possible in order to determine its response slope. In each of these cycles, 928␮l of externally prepared solutions containing 1.0×104, 5.0×104, 1.0×103or 5.0×103mol l1of silver nitrate were aspirated inside the system by means of port 4 of the SeV. Afterwards these plugs were sent to the detection system with a flow rate of 2.6 ml min1selecting port 8 of SeV and using the peristaltic pump in the for- ward pumping mode. The titration cycle was performed by

(4)

selecting port 3 and filling SoV active port channel with the volumetric solution and subsequently sending it to waste.

Volumetric solution aliquots intercalated with sample aliquots were inserted into the HC by means of on/off cycles of SoV2 when a set of electric pulses was sent to SoV. To accomplish this, small aliquots of the volumetric solution (silver nitrate) were aspirated to the HC by actuat- ing SoV2 (SoV2 on) and intercalated with aliquots of the sample solution (SoV2off). The mutual dispersion between the aliquots of both solutions throughout aspiration to the HC and sending of both zones to the detector enabled the titration to be carried out. To accomplish the titration pro- cedure, the time interval to switch SoV2 on was set to 5, 10, 15, 20, 25, 30, 35, 40, 45, 50, 55, 60, 65, 70 and 75%

of each SoV2on/off cycle period (t1).

Before sample injection, calibration with standard solu- tions of different silver concentrations was carried out. For very low concentration of volumetric solution and sample, it was necessary to determine the slope of the electrode re- sponse under the same conditions as those used during the titration program (Table 1, steps 3, 4 and 5), with volumetric solution at the active port of SoV2and carrier at the passive port of SoV2.

Some preliminary experiments were performed to estab- lish the flow injection parameters to evaluate the tubular electrode characteristics. Determinations of chloride and iodide were also carried out by conventional potentiomet- ric titration using conventional silver-selective electrode as indicator and silver nitrate as volumetric solution. This determination was performed with an automated batch titrator. The equivalence point was determined using the first derivative of the titration curves. The potentiometric titration was performed with variable volume additions of volumetric solution ranging from 0.001 to 2.0 ml, using con- centrations of volumetric solution varying from 5.0×105 to 103mol l1. The initial addition of volumetric solution was set to 0.005 ml. Volumes of analyte solutions were se- lected between 5.0 and 15.0 ml. The time period of constant potential before the next addition was set at 5 s.

3. Results and discussion 3.1. Set-up optimisation

The described SIA set-up was optimised, assessing the influence of the hydrodynamic and chemical parameters on the peak height, reproducibility and accuracy of the results.

After initial trials for selection of approximate values for each parameter, optimisation of various experimental con- ditions of the flow system, such as injection volumes, flow rate, sampling rate and switching time of SoV2, were inves- tigated and established using an univariate search approach.

This study was initially started with 5×103mol l1 solution of silver ions. Then the flow rate was assessed in the interval 2.6–5.9 ml min1and it was found that higher

sensitivity was obtained with lower flow rates, although the sampling rate suffered from a great decrease. A flow rate of 2.6 ml min1 was selected for the measurement step, under which conditions maximal response was attained.

Thereafter, several experiments were performed with as- piration flow rates of 1.46, 1.84, 2.03, 2.41, 2.6, 2.78 or 2.97 ml min1 and SoV2 cycle periods (t1) of 2.0, 3.0 or 4.0 s. Finally t1 was fixed at 4.0 s and the aspiration flow rate was maintained at 2.78 ml min1.

Having optimised the flow and sampling rate, the effect of injection volume was determined. The effect of sample volume was investigated setting different sampling times and loading an aliquot of an externally prepared silver standard solution of 5×103mol l1 into the HC (Table 1, step 1) and directing it towards the detector (Table 1, step 2). These two steps were repeated but with larger sampling intervals (9–50 s) with various increments (7 and 10 s, respectively).

By increasing the injection volume up to 927␮l (corre- sponding to sampling time of 20 s) the analytical signal of silver ions increased. The sampling time was set to 20 s, al- lowing the resulting signal to correspond to about 95% of steady-state. Using a larger volume (longer sampling time) would produce an undesirable increase in time to return the measured signal to the baseline and only a slight increase in the signal. Smaller volumes produced less reproducible signals.

Experiments to check the overall behaviour involved running a set of externally prepared silver calibration solu- tions and setting the sampling time interval at 20 s, the flow rate at 2.6 ml min1, the sampling rate at 2.78 ml min1 and the time of the switching period t1 at 4.0 s. Once the optimal conditions were established, a calibration graph for solutions ranging between 2.5×106and 0.1 mol l1 was processed to evaluate the linear and non-linear response zones. The calibration graph for the silver-selective tubular electrode was linear between 1.5×105 and 0.1 mol l1 Ag(I) in the injected solution.

3.2. Implementation of the simultaneous titration procedure

Both chloride and iodide ions promote the quantitative reaction with Ag(I) ion, resulting in the formation of a pre- cipitate under stoichiometric conditions. Since the solubility products (KSP) of the resulting precipitates are quite differ- ent, 5.0×1010mol2l2for AgCl and 2.0×1016mol2l2 for AgI in 0.1 mol l1 ionic strength adjusted media, it should be possible to have two well-defined equivalence points in the titration curves using either conventional batch equipment or the developed set-up (Fig. 2). This aspect enables the sequential determination of chloride and iodide by potentiometry using a crystalline membrane Ag(I)/Ag2S electrode.

To evaluate the minimum concentration of chloride and iodide that could be determined (taking into account the lower limit of linear response of the electrode) both the first

(5)

Fig. 2. Plots of the titration curves obtained for solutions containing both chloride and iodide. (A) Using the conventional batch titrator; (B) using the developed system. (1) and (2) refers to the equivalence points for iodide and chloride, respectively.

Table 2

Analytical ranges in which it is possible to determine chloride or iodide in pure solutions by the proposed procedure

Silver nitrate

concentration (mol l1)

Analyte concentrations limits

5.0×10−5 6.0×10−6 to 1.0×10−4

1.0×10−4 7.5×10−6 to 2.0×10−4

2.5×10−4 1.5×10−5 to 5.5×10−4

5.0×10−4 3.0×10−5 to 1.0×10−3

derivative method and the Gran’s plot method were com- pared to assess the equivalence points. In the latter method, the accuracy of the results is limited by the accurate knowl- edge of the response slope of the detection unit used.

Hence, the attainment of the actual slope value requires the calibration of the electrode under the same conditions as those used in the titration procedure. Therefore, the SoV2 passive channel was filled with the carrier solution and the active channel with the cation solution. To implement the calibration procedure the control software of the sys-

Table 3

Results obtained (mol l1) by the proposed procedure using the Gran’s plot method and by the conventional first derivative method after titration of prepared sample solutions

Analyte Titrant Sample SIA system Conventional procedure

I 5.0×105 7.5×106 7.37 (±0.13)×106 6.54 (±0.15)×106

1.0×104 7.48 (±0.12)×106 6.53 (±0.15)×106

1.2×104 7.52 (±0.15)×106 6.75 (±0.14)×106

5.0×104 3.0×105 2.93 (±0.10)×105 2.91 (±0.12)×105

Cl 5.0×10−5 5.0×10−6 4.46 (±0.12)×10−6 Not possible

7.5×10−6 7.50 (±0.12)×10−6 Not possible

8.0×10−5 1.0×10−5 1.01 (±0.10)×10−6 Not possible

5.0×10−4 1.0×10−4 9.97 (±0.14)×10−5 9.93 (±0.12)×10−5

I/Cl 1.0×104 I 5.0×106 4.92 (±0.10)×106 3.76 (±0.12)×106

Cl 3.2×105 3.11 (±0.13)×105 Not possible

4.5×10−3 I 5.0×10−6 4.92 (±0.11)×10−6 3.76 (±0.10)×10−6

Cl 3.2×10−5 3.11 (±0.11)×10−5 Not possible

Fig. 3. Gran’s plots obtained with the developed set-up for the titration of a 1×105mol l1 iodide with (a) 5×105, (b) 1×104 and (c) 1.2×104mol l1 silver nitrate solution.

tem was changed to the sequence of ports 3 and 8; this would be repeated twice automatically (Table 1, steps 3, 4 and 5).

To accomplish a halide titration the concentration of sil- ver ions in solution before the equivalence point could be determined considering both the concentration of the an- ion and the solubility product of the precipitate ([Ag+] = KSP/([X]−[Ag+added])). Concentrations are considered in- stead of activities once the ionic strength is constant. For iodide and chloride it will correspond to concentrations of free silver below the lower limit of linear response. After the equivalence point and using the equation [Ag+free] = [Ag+added]−[AgX] it is possible to calculate the concen- tration of unbound silver, which determines the analytical signal.

Table 2 shows the analytical ranges in which both ions could be determined for different concentrations of the volumetric solution. Lower limits correspond to the lowest concentration of analyte solution that are needed to obtain an equivalence point using the proposed procedure. Higher

(6)

limits were based on the assumption that a minimum of four experimental results after the equivalence point are neces- sary to estimate them. The accuracy and precision achieved in the titration of different solutions of iodide and chloride is evidenced inTable 3. As can be seen, the increase in vol- umetric solution (titrant) concentration corresponds to an increase in accuracy of the achieved results together with the increased number of experimental points which define a linear Gran’s plot (Fig. 3).

For the same conditions, conventional titrations provided inaccurate results for iodide determinations because the ef- fect of the curvilinear region of electrode response damps the potential shift of the titration curve. In a chloride titra- tion, this effect is more pronounced due to the greater sol- ubility of the produced silver chloride. For this reason it is impossible to get accurate results.

In the case of the simultaneous titration, an additional difficulty arises since the excess of silver ions after the first equivalence point should be equal to or smaller than the value determined by the chloride precipitation equilibrium in order to provide unbiased results. As can be seen inTable 3 this condition is achieved when the concentrations of both ions are low and do not differ from each other more than one decade.

Several experiments were performed with aluminium nitrate in the sample solutions in the range of 5×106 to 104mol l1 in order to evaluate its influence on co- precipitation. It was found that for low concentrations of iodide and chloride (106to 105mol l1) its influence on the quality of the results was not significant.

4. Conclusions

The proposed SIA procedure enables low concentrations of chloride and iodide in mixtures to be determined using the Gran’s plot method. This is not possible when the con- ventional procedure is used. The results showed that there are limitations to the Gran’s plot method application for analysis of mixtures of chloride and iodide when their con- centrations increase or differ significantly. In comparison with a SIA titration, determination of low concentrations

of chloride or iodide in pure solutions was inaccurate or impossible using the conventional batch titration.

Acknowledgements

The authors thank the Universidad del Pa´ıs Vasco, for its financial support for the investigation project UPV-E- 14795/2002 (Spain) and the Grant Agency of the Charles University, grant no. 250/2002/B-CH/FaF (Czech Republic).

References

[1] W.J. Blaedel, R.H. Laessig, Anal. Chem. 36 (1964) 1617.

[2] W.J. Blaedel, R.H. Laessig, Anal. Chem. 37 (1965) 332.

[3] J.M. Calatayud, P.C. Falcó, R.M. Abert, Analyst 112 (1987) 1063.

[4] J. Ruzicka, E.H. Hansen, Anal. Chim. Acta 78 (1975) 145.

[5] M. Korn, L.F.B.P. Gouveia, E. Oliveira, B.F. Reis, Anal. Chim. Acta 313 (1995) 177.

[6] R.N. Fernandes, M.G.F. Sales, B.F. Reis, E.A.G. Zagatto, A.N.

Araújo, M.C.B.S.M. Montenegro, J. Pharm. Biomed. Anal. 25 (2001) 713.

[7] E.P. Serjeant, in: P.J. Elving, J.D. Winefordner (Eds.), Potentiometry and Potentiometric Titrations, Wiley, New York, 1984, p. 465.

[8] A.M. Pimenta, N.A. Araújo, M.C.B.S.M. Montenegro, Anal. Chim.

Acta 438 (2001) 31.

[9] S.S.M. Hassan, S.A.M. Marzouk, A.M. Sayed, Electroanalysis 855 (1993) 5.

[10] I.M.P.L.V.O. Ferreira, J.L.F.C. Lima, A.S.S. Rangel, Rev. Port. Farm 45 (1995) 24.

[11] I.M.P.L.V.O. Ferreira, J.L.F.C. Lima, A.S.S. Rangel, Anal. Sci. 10 (1994) 801.

[12] A. Andrade-Eiroa, J.A. Erustes, R. Forteza, V. Cerda, J.L.F.C. Lima 467 (2002) 25.

[13] J. Alpizar, A. Crespi, A. Cladera, R. Forteza, V. Cerda, Electroanal- ysis 8 (1996) 1051.

[14] J. Alpizar, A. Crespi, A. Cladera, R. Forteza, V. Cerda, Lab. Rob.

Autom. 8 (1996) 165.

[15] I.M.P.L.V.O. Ferreira, J.L.F.C. Lima, L.S.M. Rocha, Fresenius J.

Anal. Chem. 347 (1993) 314.

[16] A.N. Araújo, R.C.C. Costa, J.L.F.C. Lima, B.F. Reis, Anal. Chim.

Acta 358 (1998) 111.

[17] S. Alegret, J. Alonso, J. Bartroli, J.M. Paulis, J.L.F.C. Lima, A.A.S.C.

Machado, Anal. Chim. Acta 164 (1984) 147.

[18] S. Alegret, J. Alonso, J. Bartroli, A.A.S.C. Machado, J.L.F.C. Lima, J.M. Paulis, Quim. Anal. 6 (1987) 278.

Referências

Documentos relacionados

S UMÁRIO Apresentação 6 1 Violência, crime e políticas de segurança pública no Brasil contemporâneo 13 Luís Antônio Francisco de Souza 2 O sistema policial no estado de São Paulo