Eletrodos Externos
para Biopotenciais
Prof. Sérgio F. Pichorim
Figure 5.1 Electrode-electrolyte interface The current crosses it from left to
© From J. G. Webster (ed.), Medical instrumentation: application and design. 3rded. New York: John Wiley & Sons, 1998.
• C ÍÎ C+ +
• A- ÍÎ A +
e-• Metal Î Solução : Predomina a Oxidação
• Solução Î Metal : Predomina a Redução
• Interface Metal-Solução : Gradiente de
Potencial de Meia-célula
• Uma meia-célula é uma estrutura que
contém um eletrodo condutor rodeado de
um eletrólito condutivo. Reações químicas
nesta interface bombeiam cargas elétricas
entre o eletrodo e o eletrólito, resultando
em uma d.d.p.
• Este potencial não pode ser medido
diretamente: Faz-se referência com o gás
hidrogênio.
© From J. G. Webster (ed.), Medical instrumentation: application and design. 3rded. New York: John Wiley & Sons, 1998.
• Exemplo: Zn
2+
= -0,76V e Cu
+
= +0,52V
Eletrodo com Corrente Elétrica
• Surgimento de Sobre potencial (Vsp)
• Vsp = Vohm + Vcon + Vativ + Vjunc
• Efeito ôhmico : resistência não-linear do eletrólito
• Variação da Concentração
• Atividade iônica: diferença de energia entre a
redução e a oxidação
© From J. G. Webster (ed.), Medical instrumentation: application and design. 3rded. New York: John Wiley & Sons, 1998.
• Eletrodo perfeitamente polarizável Î não
há transferência de carga (totalmente
capacitivo) Î Metais nobres são
polarizáveis (difícil oxidação e redução)
• Eletrodo perfeitamente não-polarizável Î
as cargas fluem livremente (totalmente
resistivo ou curto-circuito). Um bom
eletrodo não-polarizável Î Ag/AgCl
Figure 5.4 Equivalent circuit.
CdÎ Cargas entre Eletrodo e Eletrólito
RdÎ Fuga destas Cargas
© From J. G. Webster (ed.), Medical instrumentation: application and design. 3
Figure 5.6 Experimentally determined magnitude of impedance as a function
rded. New York: John Wiley & Sons, 1998.© From J. G. Webster (ed.), Medical instrumentation: application and design. 3rded. New York: John Wiley & Sons, 1998.
• AgCl é pouco solúvel em água
• Ag ÅÆ Ag+ +
e-• Ag+ + Cl- ÅÆ AgCl
(deposita-se no eletrodo)
• A atividade do Ag+ e Cl- é baixa
• Vativ = 0
1 - Eletrodo sintetizado de Ag/AgCl. Pó comprimido e aquecido.
© From J. G. Webster (ed.), Medical instrumentation: application and design. 3rded. New York: John Wiley & Sons, 1998.
2 - Camada de AgCl depositado eletroliticamente.
Espessura depende da corrente e do tempo.
Resistência depende da espessura.
Construção:
Ag AgCl
Cl-Figure 5.5 Impedance as a function of frequency for Ag electrodes coated with an
electrolytically deposited AgCl layer. The electrode area is 0.25 cm
2. Numbers attached to
curves indicate the number of mA⋅s for each deposit. (From L. A. Gedders, L. E. Baker, and
A. G. Moore, "Optimum Electrolytic Chloriding of Silver Electrodes," Medical and
© From J. G. Webster (ed.), Medical instrumentation: application and design. 3rded. New York: John Wiley & Sons, 1998.
Figure 10.2 pH electrode (From R. Hicks, J. R. Schenken, and M. A. Steinrauf,
Laboratory Instrumentation. Hagerstown, MD: Harper & Row, 1974. Used with
permission of C. A. McWhorter.)
Figure 5.7 Magnified section of skin, showing the various layers (Copyright ©
1977 by The Institute of Electrical and Electronics Engineers. Reprinted with
permission, from IEEE Trans. Biomed. Eng., March 1977, vol. BME-24, no. 2, pp.
Queratina
Células mortas
Novas células
Eletricamente
similar ao resto
do corpo.
© From J. G. Webster (ed.), Medical instrumentation: application and design. 3rded. New York: John Wiley & Sons, 1998.
Figure 5.8 A body-surface electrode is placed against skin, showing the total
electrical equivalent circuit obtained in this situation. Each circuit element on the
right is at approximately the same level at which the physical process that it
represents would be in the left-hand diagram.
Sweat glands
and ducts
Electrode
Epidermis
Dermis and
subcutaneous layer
R
uR
eE
seE
heR
sR
dC
dE
PR
PC
PC
eGel
Z pele (1 cm
2):
200 k
Ω a 1 Hz
200
Ω a 1 MHz
• Artefato de Movimento Î Em eletrodos
polarizáveis, o movimento mecânico entre
eletrodo e eletrólito reorganiza as cargas
em Cd (gerando tensões Vcd).
• Mudanças nos potenciais e geração de
ruído de baixa freqüência.
• Aço inox, Platina, Liga de prata e níquel
(polarizáveis) necessitam boa aderência!
© From J. G. Webster (ed.), Medical instrumentation: application and design. 3rded. New York: John Wiley & Sons, 1998.
Figure 5.9 Body-surface biopotential electrodes (a) Metal-plate electrode used
for application to limbs. (b) Metal-disk electrode applied with surgical tape. (c)
Disposable foam-pad electrodes, often used with electrocardiograph monitoring
apparatus.
Figure 5.10 A metallic suction electrode is often used as a precordial electrode on
clinical electrocardiographs.
Área de contato menor Î Maior Impedância!
Utilização por pouco tempo.
© From J. G. Webster (ed.), Medical instrumentation: application and design. 3rded. New York: John Wiley & Sons, 1998.
Aplicação do Eletrodo
• Limpeza da pele
• Remoção de pelos
• Esfoliação
• Gel eletrólito
• Evitar curto (via gel) entre os eletrodos
próximos
Outros tipos de Eletrodos
• Eletrodo flutuante Î diminui artefatos
• Eletrodo flexível Î maior conforto
• Eletrodo para neonatos (pele sensível)
• Eletrodo não metálicos Î transparente ao
raio-X
© From J. G. Webster (ed.), Medical instrumentation: application and design. 3rded. New York: John Wiley & Sons, 1998.