Half Life of 127 Te

Wagner F. Batista, Frederico A. Genezini, Cibele B. Zamboni, Guilherme S. Zahn

Centro do Reator de Pesquisas (CRPq) - Instituto de Pesquisas Energéticas e Nucleares (IPEN - CNEN/SP). Av. Linneu Prestes, 2242 – Cidade Universitária, São Paulo, SP, 05507-000, Brazil

Abstract. In this work, the half life of the beta-unstable nucleus

127

Te was studied using neutron- irradiated samples of

126

Te. The gamma activity of each of the irradiated samples was followed for 3-5 consecutive half lives. The results were analysed in two different ways, and the resulting half-life was 9.295(5)h, which is compatible with the tabulated value of 9.35(7)h, with much lower uncertainty.

Keywords:

127

Te; half life PACS: 21.10.Tg; 23.40.-s; 27.60.+j

INTRODUCTION

The half life is a very important parameter on nuclear decays; in particular, the half life of

127

Te is of relevance both to stellar evolution simulations [1] and neutron activation analysis [2]. The presently accepted value [3], is 9.35(7)h, which presents a relative standard deviation of 0.7% and a decrease in this uncertainty would reflect directly in both applications.

EXPERIMENTAL PROCEDURE

In the present experiment, the

127

Te samples were produced by neutron irradiation of 10mg-

126

Te-enriched samples for 6 minutes in the pneumatic station of the IEA-R1 nuclear reactor under a thermal neutron flux of approximately 5x10

12

cm

-2

s

-1

. A total of 13 radioactive samples were produced and counted on a 60%-HPGe detector coupled to a 4096-channel MCA; each sample was counted for approximately 40 separate consecutive 1h acquisitions in order to allow for the decay analysis.

CP1139, Nuclear Physics 2008, XXXI Workshop on Nuclear Physics in Brazil, edited by V. Guimaraes, J. R. B. Oliveira, K. C. D. Macario, and F. A. Genezini

© 2009 American Institute of Physics 978-0-7354-0676-6/09/$25.00

187

(2)

DATA ANALYSIS

For the determination of the half-life of

127

Te, the 417.9keV peak (I

γ

=100%) was fitted to a gaussian in each spectrum and the peak area (C) was fitted, using a covariant Gauss-Marquardt routine implemented in MatLab environment, to a simple exponential decay (Eq.1) by two different approaches:

• In the first, C(t) was determined individually for each sample and then the weighted average of the values obtained for all samples was calculated (Fig. 1a);

• In the second, data from all samples were gathered in a single fit together with a “normalization” parameter for each, so that a single value for the half-life is fitted (Fig. 1b).

T1B = 9.2950 (8) h

total counUng lime: 431 h

(1)

200 300 Time (hours)

FIGURE 1. Left (1a): Individual fits of the 13 samples;

Right (1b): All 43 1h of data acquisition fitted in a single session;

RESULTS

The weighted average of the 13 individual fits resulted in a half-life of 9.294(5)h, while the result of the complete 431h fit was 9.295(5)h; both values are almost identical, and both agree with the value found in reference [3] (9.35(7)h), with enhancement of more than 1 order of magnitude in the uncertainty.

REFERENCES

1. R. Reifarth and F. Käppeler, Phys. Rev. C, 66, 054605 (2002).

2. T. C. Hughes, J. Radioanal. Nucl. Ch. 59, 7-13 (1980).

3. K. Kiato and M. Oshima, Nucl. Data sheets 77, 1-124 (1996).

188

(3)

Referências

Documentos relacionados

Ao Dr Oliver Duenisch pelos contatos feitos e orientação de língua estrangeira Ao Dr Agenor Maccari pela ajuda na viabilização da área do experimento de campo Ao Dr Rudi Arno

Neste trabalho o objetivo central foi a ampliação e adequação do procedimento e programa computacional baseado no programa comercial MSC.PATRAN, para a geração automática de modelos

Ousasse apontar algumas hipóteses para a solução desse problema público a partir do exposto dos autores usados como base para fundamentação teórica, da análise dos dados

didático e resolva as ​listas de exercícios (disponíveis no ​Classroom​) referentes às obras de Carlos Drummond de Andrade, João Guimarães Rosa, Machado de Assis,

i) A condutividade da matriz vítrea diminui com o aumento do tempo de tratamento térmico (Fig.. 241 pequena quantidade de cristais existentes na amostra já provoca um efeito

The probability of attending school four our group of interest in this region increased by 6.5 percentage points after the expansion of the Bolsa Família program in 2007 and

If, on the contrary, our teaching becomes a political positioning on a certain content and not the event that has been recorded – evidently even with partiality, since the

Alguns ensaios desse tipo de modelos têm sido tentados, tendo conduzido lentamente à compreensão das alterações mentais (ou psicológicas) experienciadas pelos doentes