• Nenhum resultado encontrado

Flow of physiological fluids in microchannels: the sedimentation effect

N/A
N/A
Protected

Academic year: 2021

Share "Flow of physiological fluids in microchannels: the sedimentation effect"

Copied!
8
0
0

Texto

(1)

IFMBE Proceedings

Volume 31

Series Editor: R. Magjarevic

Jointly Organised by

Biomedical Engineering Society (Singapore)

Global Enterprise for Micromechanics and Molecular

Medicine (GEM4)

National University of Singapore

(2)

IFMBE Proceedings Vol. 31

C.T. Lim •

J.C.H. Goh (Eds.)

6th World Congress of Biomechanics

(WCB 2010)

August 1–6, 2010

Singapore

In Conjunction with 14th International Conference on Biomedical

Engineering (ICBME) and 5th Asia Pacific Conference on Biomechanics

(APBiomech)

(3)

Editors

C.T. Lim

National University of Singapore Fac. Engineering

Dept. Mechanical Engineering Div. Bioengineering Engineering Drive 7 117574 Singapore 1 Block E3A #04-15 Singapore Email: ctlim@nus.edu.sg J.C.H. Goh

National University of Singapore Dept. Orthopaedic Surgery Tissue Engineering Program Medical Drive 27

117510 Singapore

Level 4, DSO (Kent Ridge) Bldg. Singapore

Email: dosgohj@nus.edu.sg

ISSN 1680-0737

ISBN 978-3-642-14514-8 e-ISBN 978-3-642-14515-5 DOI 10.1007/978-3-642-14515-5

Library of Congress Control Number: Applied for

© International Federation for Medical and Biological Engineering 2010

This work is subject to copyright. All rights are reserved, whether the whole or part of the material is concerned, specifically the rights of translation, reprinting, reuse of illustrations, recitation, broadcasting, reproduction on microfilm or in any other way, and storage in data banks. Duplication of this publication or parts thereof is permitted only under the provisions of the German Copyright Law of September 9, 1965, in its current version, and permis-sions for use must always be obtained from Springer. Violations are liable to prosecution under the German Copyright Law.

The use of general descriptive names, registered names, trademarks, etc. in this publication does not imply, even in the absence of a specific statement, that such names are exempt from the relevant protective laws and regulations and therefore free for general use.

The IFMBE Proceedings is an Official Publication of the International Federation for Medical and Biological Engineering (IFMBE)

Typesetting: Scientific Publishing Services Pvt. Ltd., Chennai, India.

Cover Design: deblik, Berlin

Printed on acid-free paper 9 8 7 6 5 4 3 2 1 springer.com

(4)

Table of Contents LVII

Comparison of Effects of Various Methods of Recovery of Muscle after Applied Exercise . . . 1040

P. Nov´akov´a, P. ˇSifta

Theme 4: Cell Mechanics

Flow Behaviour of Labeled Red Blood Cells in Microchannels: A Confocal Micro-PTV

Assessment . . . 1047

R. Lima

Fractal and Image Analysis of Cytoskeletal F-Actin Orgnization in Endothelial Cells under

Shear Stress and Rho-GDIα Knock Down . . . 1051

Ying-Xin Qi, Xiao-Dong Wang, Ping Zhang, Zong-Lai Jiang

Swirling Flow Can Suppress Platelet Adhesion to the Surface of a Sudden Tubular Expansion Tube . . . 1055

F. Zhan, Y.B. Fan, X.Y. Deng

Numerical Study on Effects of Liposome-Encapsulated Hemoglobin on Blood Flows at

Microvascular Bifurcation with Considering Erythrocyte Aggregation . . . 1059

T. Hyakutake, Y. Akagi, T. Imaru, T. Matsumoto, S. Yanase

Wall Shear Stress Distribution Inside Induced Cerebral Aneurysm on Rabbit . . . 1063

T. Tanoue, S. Tateshima, D. Wakui, F. Vinuela, R. Sudo, K. Tanishita

Red Blood Cell Dispersion in 100 μm Glass Capillaries: The Temperature Effect . . . 1067

D. Pinho, A. Pereira, R. Lima, T. Ishikawa, Y. Imai, T. Yamaguchi

Flow of Physiological Fluids in Microchannels: The Sedimentation Effect . . . 1071

V. Garcia, T. Correia, R. Dias, R. Lima

Drift and Fluctuating Motion of Artificial Platelet during Adhesion Process Near the Wall . . . 1075

H. Tobimatsu, A. Paragon, Y. Okamura, S. Takeoka, R. Sudo, K. Tanishita

Simultaneous Topography and Elasticity Measurement of Live PC-12 Cells by Using

Amplitude-Modulation Atomic Force Microscopy . . . 1079

M.C. Liu, S. Tien, C.-C.K. Lin, M.-S. Ju

Numerical Simulations of Vesicular Driving Forces Inside Living Cells . . . 1083

D. Robert, C. Wilhelm

Effect of Cyclic Stretch on the Visco-Elastic Deformation of Endothelial Cells in Micropipette Aspiration Experiment . . . 1087

Javad Hatami, Mohammad Tafazzoli-Shadpour, Mohammad Ali Shokrgozar

Strain Magnitude and Strain Rate Influence Stretch-Induced Injury of PC12 Cells . . . 1091

H. Nakadate, S. Aomura, Y. Zhang, A. Kakuta, S. Fujiwara

Effect of Extracellular Matrix Stiffness on Ductular Formation of Biliary Epithelial Cells . . . 1095

Tomoya Komatsu, Ryo Sudo, Toshihiro Mitaka, Mariko Ikeda, Kazuo Tanishita

Receptor-Ligand Bond Spacing and Stresses in Membrane Bulge of Cell Adhesion . . . 1099

(5)

Flow of physiological fluids in microchannels: the sedimentation effect

V. Garcia

1

, T. Correia

1, 2

, R. Dias

1, 3

, R. Lima

1, 3

1

Instituto Politécnico de Bragança (IPB), Campus Stª Apolónia, Apartado 134, 5301-857 Bragança, Portugal

2

CIMO, ESA, Instituto Politécnico de Bragança (IPB), Campus Stª Apolónia, Apartado 134, 5301-857 Bragança, Portugal.

3

CEFT - Centro de Estudos de Fenómenos de Transporte, Faculdade de Engenharia da Universidade do Porto, 4200-465 Porto, Portugal.

Abstract— Microfluidic devices are becoming one of the

most promising new tools for diagnostic applications and treatment of several chronic diseases. Hence, it is increasingly important to investigate the rheological behaviour of physio-logical fluids in microchannels. The main purpose of the pre-sent experimental work is to investigate the flow of two differ-ent physiological fluids frequdiffer-ently used in microfluidic devices. The working fluids were physiological saline (PS) and dextran 40 (Dx40) containing about 6% of sheep red blood cells (RBCs), respectively. The capillaries were placed horizontally on a slide glass and the flow rate of the working fluids was kept constant by using a syringe pump. By means of a camera the images were taken and transferred to the computer to be ana-lysed. Generally, the results show that PS and Dx40 have dif-ferent flow behaviour due to the sedimentation of the RBCs.

Keywords— Physiological fluids, Red blood cells,

microflu-idic devices, sedimentation, microcirculation.

I.

I

NTRODUCTION

In microcirculation there are two well known phenom-ena, i. e., the Faharaeus effect and Faharaeus-Lindqvist effect. These two effects are strongly related to the micro-tube diameter and other physical and hemorheological fac-tors. An explanation for the Faharaeus-Lindqvist effect is the decrease of the dynamic hematocrit (Hct) with decreas-ing microtube diameter. This effect can be explained by the formation of a cell-free layer (CFL) near the wall of the microtube. This CFL located between the RBC core and wall contributes to the reduction of apparent blood viscos-ity. However, the complex formation of the CFL has not yet convincingly demonstrated mainly due to multi-physical and hemorheological factors that affect CFL [1-3].

Currently, biomedical microdevices are becoming one of the most promising tools for the diagnostic and treatment of several diseases, such as diabetes, malaria and cancer. Hence, it is increasingly important to investigate the rheological behaviour of physiological fluids in microchan-nels in order to make use on the physics of microfluidics to either develop new lab-on-chip devices or to optimize the design of the existent microfluidic chips.

In this paper we investigated the flow behaviour of two different physiological fluids frequently used in biomedical microdevices. The working fluids used in this study were

physiological saline (PS) and dextran 40 (Dx40) containing about 6% of sheep red blood cells (RBCs), respectively. By using a syringe pump and a camera it was possible to meas-ure qualitatively flow behaviour within a horizontal capil-lary.

II.

M

ATERIALS & METHODS

Blood sample preparation

Four working fluids were used in this study: physiological saline (PS) containing about 6% (6Hct) of sheep red blood cells (RBCs), dextran 40 (Dx40; Otsuka Medicine) contain-ing about 6% (6Hct) of sheep RBCs, PS with 10% of dex-tran 40 (Dx40s10; Sigma aldrich, BioChemika)and PS con-taining 30% of dextran 40 (Dx40s30; Sigma aldrich, BioChemika) containing in both fluids about 2% (3Hct) of sheep RBCs.

The blood was collected from a healthy adult sheep, where heparin was added to prevent coagulation. The RBCs where separated from the bulk blood by centrifugation (3000 RPM for 15 min) and aspiration of the plasma and buffy coat and then washed twice with PS. The washed RBCs where diluted with PS to make up the required RBCs concentration by volume. The hematocrit (Hct) of the RBCs suspension sample was about 6% (6Hct). Note that all the blood samples were stored hermetical at 4ºC until the ex-periment was performed at room temperature (18 to 20ºC). Figure 1 shows the main steps for the preparation of the blood samples.

(6)

Experimental setup

To investigate the RBC sedimentation we have decided to use two kinds of experimental methods. In the first method the capillaries were placed horizontally on a slide glass and by using a syringe pump (New Era Pump Sys-tems, USA) a pressure-driven flow was kept constant at 50 l/min which corresponds to a Reynolds ~0.9(PS) ~0.3(Dx40) (see Fig.2). For the second method we decided to fill 4 tubes with different kinds of physiological fluids containing sheep RBCs, i.e., PS, Dx40, Dx40s10, Dx40s30 (see Fig.3). Additionally the visualization of the flow in microchannels was possible by means of a high-speed video microscopy system (see Fig.4). Detailed information about this latter microvisualization system can be found elsewhere [4-6].

Fig. 2 Experimental setup for the dynamic sedimentation measurements.

Fig. 3 Experimental setup for the static sedimentation measurements.

Fig. 4 Experimental setup for the visualization of the in vitro blood flow in glass microchannels (adapted from 4, 6).

III.

R

ESULTS & DISCUSSION

To analyse the dynamic sedimentation of PS and Dx40 containing RBCs we decided to use flow rates close to the one observed in vivo, i.e., 10 l/min. During the experiment we made flow qualitative visualizations measurements in glass capillaries with diameters of about 1.2 mm. The visu-alizations were captured by a camera for about 15 minutes. Figure 5 shows the flow qualitative measurements for 0 minutes and 15 minutes. This image shows clearly that for a period of 15 minutes the RBC tend to settle down in the fluid with PS whereas using Dx40 we did not observe any RBC sedimentation. Although not shown in Figure 5, for the case of PS fluid we did not observe any RBC sedimenta-tion for the first 10 minutes. According to our visualizasedimenta-tion the RBCs tend to settle down for period of time superior to 10 minutes.

Fig. 5 Dynamic sedimentation measurements for two different time periods of PS and Dx40 containing RBCs (Flow rate = 50 l/min) .

(7)

To study the static sedimentation of the sheep RBCs in PS and Dx40 we decided to fill 4 tubes containing identical fluids to the ones used in the previous experiments and also additional fluids containing PS with 10% of Dextran 40 ( Dx40s10) and PS with 30% of Dextran 40 (Dx40s30). The visualizations were captured by a camera for a period of about 2 hours. Figure 6 shows a summary of the most rele-vant visualization results. This image shows that for a pe-riod of 30 minutes the RBC tend to settle down in the fluid with PS. The sedimentation tends to increase with the time. For a period of 1 hour and 30 minutes it is very clear the interface between the high concentration of cells and the cell-free fluid. For a period superior to 1 hour little sedi-mentation was observed in the sample Dx40s10. For the case of Dx40 and Dx40s30 we did not observe any signifi-cant RBC sedimentation.

Fig. 6 Static sedimentation measurements for a period of 2 hours of PS, Dx40, Dx40s10 and Dx40s10 containing RBCs .

Additionally flow visualization measurements were also performed in glass microchannels (see Fig.7) and compared with in vivo blood flow (see Fig.8). Figure 7 shows that for the case of Dx40 there is a clear formation of cell-free layer adjacent to the walls of microchannels. However, in the fluid with PS the RBCs do not exhibit a clear tendency to migrate into the microtube axis. The in vivo visualization measurements (Fig. 8) have shown a clear tendency for the formation of a plasma layer in microvessels [6, 7]. These results indicate that in vitro blood containing Dx40 has a flow behaviour closer to the one observed in vivo microves-sels.

Fig. 7 In vitro flow visualization in glass microchannels for a period time bigger than 10 minutes a) Dx40 containing RBCs; b) PS containing RBCs.

(8)

IV.

C

ONCLUSION & FUTURE DIRECTIONS

The flow behaviour of two different physiological fluids frequently used in biomedical microdevices was investi-gated. Our preliminary results indicate that in vitro blood containing Dx40 has a flow behaviour closer to the one observed in vivo microvesseles. The in vitro blood contain-ing PS did not show a clear formation of cell-free layer which might be due to the fast sedimentation of the RBCs. In the near future we plan to vary the flow rate and diameter to study the influence of these effects on the RBC sedimen-tation.

A

CKNOWLEDGMENT

This study was supported in part by the following grants: Grant-in-Aid for Science and Technology (PTDC/SAU-BEB/108728/2008 and PTDC/SAU-BEB/105650/2008) from the Science and Technology Foundation (FCT) and COMPETE, Portugal. The authors would like to thank the students from the Master of Biomedical Technology, ES-TiG, IPB, for their valuable technical assistance in this research work.

R

EFERENCES

1. Lima R, Ishikawa T, et al. (2010) Blood flow behavior in

microchan-nels: advances and future trends. Single and two-Phase Flows on Chemical and Biomedical Engineering. Bentham (in press).

2. Goldsmith H, Turitto V (1986) Rheological aspects of thrombosis and

haemostasis: basic principles and applications. ICTH-Report-Subcommittee on Rheology of the International Committee on Thrombosis and Haemostasis. Thromb Haemost. 55(3): 415–435

3. Maeda N (1996) Erythrocyte rheology in microcirculation. Japanese

Journal of Physiology 46, 1-14

4. Lima R, Wada S, Tsubota K, Yamaguchi T (2006) Confocal

micro-PIV measurements of three dimensional profiles of cell suspension flow in a square microchannel. Measurement Science and Technology 17: 797-808

5. Lima R, et al. (2009) Measurement of individual red blood cell

mo-tions under high hematocrit condimo-tions using a confocal micro-PTV system, Annals of Biomedical Engineering, 37, 1546-1559

6. Kim S, Kong, R L, et al (2006) A computer-based method for

deter-mination of the cell-free layer with in microcirculation. Microcircula-tion 13: 199-207.

7. Minamiyama M (2000) In Vivo Microcirculatory Studies at

http://www.ne.jp/asahi/minamiya/medicine/

Author: Valdemar Garcia

Institute: Instituto Politécnico de Bragança

Street: Campus Santa Apolónia

City: Braganca

Country: Portugal

Email: valdemar@ipb.pt

Referências

Documentos relacionados

Neste trabalho o objetivo central foi a ampliação e adequação do procedimento e programa computacional baseado no programa comercial MSC.PATRAN, para a geração automática de modelos

To understand how Taylor bubbles flow in microcirculation, in particular, how bubbles disturb the blood flow at the scale of blood cells, experiments were performed in microchannels

Within this, it is proposed that the learning motivations positively influences the sense dimension of solo women travel experiences because predispose women to

Dos professores participantes da pesquisa, a maioria afirma que teve em sua formação superior um componente curricular que tratou do tema avaliação (Prof.1, Prof. 5) – apenas

Quando há planejamento adequado e participação da comunidade local durante o processo de preparação de uma região para o desenvolvimento

O ensino a distância da Universidade do Norte do Paraná (UNOPAR) através do Sistema “Colaborar” faz parte do grupo Kroton Educacional que tem como fim um modelo

Tecnologicamente temos diversos programas específicos de (TCFC), para a obtenção de mensurações lineares, portantoiremos realizar um estudo comparativo entre dois

Extinction with social support is blocked by the protein synthesis inhibitors anisomycin and rapamycin and by the inhibitor of gene expression 5,6-dichloro-1- β-